The Recall Desk

Manufacturer

Pride Mobility Products Corp.

15 recalls in our database name Pride Mobility Products Corp. as the manufacturing or recalling firm.

What the numbers show

Total
15
Critical
0
Severe
3
Most recent
2026-08-12

Of the 15 recalls from Pride Mobility Products Corp. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 3 (20%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–15 of 15

  • SevereFDA (Devices)·Z-2843-2026·2026-08-12

    Powered Wheelchair Controller Software Recall by Pride Mobility

    Pride Mobility is recalling 1,102 Q6 Edge 3 Stretto powered wheelchairs with controller software version 2.0.11 due to a defect that may cause unexpected brake stops or control system crashes.

    Product
    Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2850-2026·2026-08-12

    Quantum Rehab Q6 Edge 4 Wheelchairs Recalled Due to Controller Software Issue

    Pride Mobility Products Corp. is recalling Quantum Rehab Q6 Edge 4 powered wheelchairs because a software issue can cause unexpected braking and potential injury.

    Product
    Brand Name: Quantum Rehab Product Name: Q6 Edge 4 Model/Catalog Number: Q6 Edge 4 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2852-2026·2026-08-12

    Quantum Rehab 4Front Powered Wheelchair Controller Software Recall

    Pride Mobility is recalling 1,102 Quantum Rehab 4Front powered wheelchairs with controller software version 2.0.11 due to a defect that may cause unexpected brake stops or control system crashes.

    Product
    Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2842-2026·2026-08-12

    Quantum Outback Powered Wheelchair Controller Recalled Due to Unexpected Braking Risk

    Pride Mobility Products Corp. is recalling Quantum Outback powered wheelchairs due to a software issue that may cause unexpected braking and potential injury.

    Product
    Brand Name: Quantum Rehab Product Name: Quantum Outback Model/Catalog Number: Quantum Outback Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2849-2026·2026-08-12

    Quantum Rehab Q6 Edge 3 Powered Wheelchair Controller Recalled for Unexpected Braking

    Pride Mobility Products Corp. is recalling the Quantum Rehab Q6 Edge 3 powered wheelchair controller due to a software issue that can cause unexpected braking and potential injury.

    Product
    Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2841-2026·2026-08-12

    Powered wheelchair controller recalled for unexpected braking hazard

    Pride Mobility Products Corp. is recalling 1,102 Quantum 4Front 2 wheelchair controllers because reminder pop-ups can trigger unexpected park brake stops while driving, causing falls or injuries.

    Product
    Brand Name: Quantum Rehab Product Name: Quantum 4Front 2 Model/Catalog Number: Quantum 4Front 2 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2844-2026·2026-08-12

    Quantum Rehab R-TRAK Powered Wheelchair Controller Recalled Due to Software Issue

    Pride Mobility Products Corp. is recalling Quantum Rehab R-TRAK powered wheelchairs due to a software issue that can cause unexpected braking or system crashes.

    Product
    Brand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2846-2026·2026-08-12

    Quantum Rehab Q6 Edge 2.0 Power Wheelchair Recall

    Pride Mobility is recalling 1,102 Q6 Edge 2.0 power wheelchairs due to a software issue that may cause unexpected braking or control system crashes during use.

    Product
    Brand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2851-2026·2026-08-12

    Quantum 4Front 2HD Powered Wheelchair Recall Due to Potential Safety Issues

    A recall has been issued for the Quantum 4Front 2HD powered wheelchair because its controller software may cause unexpected braking or system failure that could lead to falls or injury.

    Product
    Brand Name: Quantum Rehab Product Name: Quantum 4Front 2HD Model/Catalog Number: Quantum 4Front 2HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2848-2026·2026-08-12

    Quantum Rehab Q6 Edge 2.0 Wheelchair Recall Due to Safety Risks

    The FDA is recalling the Quantum Rehab Q6 Edge 2.0 powered wheelchair because its controller may cause unexpected braking or system crashes that could lead to falls and injuries.

    Product
    Brand Name: Quantum Rehab Product Name: Q6 Edge 2.0 (ilevel) Model/Catalog Number: Q6 Edge 2.0 (ilevel) Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2853-2026·2026-08-12

    Quantum Rival Powered Wheelchair Controller Recall for Brake Stop Hazard

    Pride Mobility Products Corp. is recalling 1,102 Quantum Rival powered wheelchair controllers due to a software issue that can cause unexpected brake stops while driving, potentially resulting in falls or injuries.

    Product
    Brand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2847-2026·2026-08-12

    Powered Wheelchair Controller Software May Cause Unexpected Brake Stop

    Pride Mobility is recalling 1,102 Quantum 1450 powered wheelchairs with controller software version 2.0.11 because reminder pop-ups can trigger an unexpected park brake stop while driving, risking falls or injury.

    Product
    Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2845-2026·2026-08-12

    Powered Wheelchair Controller Software May Cause Unexpected Brake Stop

    Pride Mobility is recalling 1,102 Q6 Edge HD powered wheelchairs due to a software issue in the controller that may cause unexpected park brake stops or system crashes when reminder pop-ups appear during use.

    Product
    Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0942-2017·2017-01-11

    Pride Mobility Recalls TRU-Balance 3 Power Positioning Systems

    Pride Mobility Products Corp is recalling 12,343 TRU-Balance 3 Power Positioning Systems because the back and tilt base can separate when tilted beyond the specified angle.

    Product
    TRU-Balance 3 Power Positioning Systems, TB3 Lift/Tilt/Recline (Gen 1). Mobility Medical Device
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1941-2012·2012-07-11

    Pride Mobility Victory 9 Wheelchair Recalled for Harness Sealing Defect

    Pride Mobility is recalling 186 Victory 9 motorized wheelchairs because the main harness connector lacks required back shield sealing, allowing water to seep through.

    Product
    Victory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.
    Category
    Medical Device
    Distribution
    0 states

About this company

Community-sourced company facts from Wikidata, not from any recall agency. This is background on the corporate entity — it says nothing about the recall(s) listed on this page, which come from the agency source notice above.

Pride Mobility Products (United States)

company in Exeter, United States

Country
United States
Founded
1986

Official website

Source: Wikidata (Q104481384), licensed CC0. Wikidata is community-edited and may not reflect the current corporate structure.