GE SIGNA 1.5T MR Systems May Overheat Implants in Low SAR Mode
GE Medical Systems recalled 2,146 SIGNA 1.5T MR systems that may overheat implants during certain imaging sequences in Low SAR mode due to a software defect in radiofrequency safety calculations.
- Product
- 1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
- Category
- Medical Device
- Distribution
- 0 states