Blood Gas Analyzer Software Issue Causes Erroneous Creatinine Results During Startup
Nova Biomedical Prime Plus Analyzers may report incorrect creatinine levels during the first 2-4 hours after installing a new sensor cartridge. These errors may not be identified by traditional quality control practices.
- Product
- Nova Biomedical Prime Plus Analyzer -Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 57400 Stat Profile Prime Plus Analyzer (2) PN 59508 Stat Profile Prime Plus Analyzer, Remanufactured
- Category
- Medical Device
- Distribution
- Distributed nationwide