The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0354-2027·Announced 2026-11-11

Baxter Recalls TruSystem 7500 Operating Table Column Due to Lift Failure

Baxter Healthcare is recalling TruSystem 7500 Hybrid Plus operating table columns because the main lift mechanism may unexpectedly drop, posing a risk of patient injury.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, and the source describes a hazard where an uncontrolled drop of the surgical table may result in serious or critical adverse health consequences, including traumatic injury and airway compromise. The source reports three complaints but states there were no reported patient injuries.

Plain-English summary

Baxter Healthcare Corporation is recalling the TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number 1854087. The recall involves one unit with serial number 107971925 and UDI-DI number 00887761974067. The product was distributed nationwide in the United States and worldwide.

The recall is due to an issue with the main lift mechanism, which may unexpectedly drop from its set position to its lowest position during use. This issue is associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. If the drop occurs, the tabletop could not be raised again. An uncontrolled drop of the surgical table may result in abrupt patient displacement during surgery, potentially causing serious or critical adverse health consequences, including traumatic injury, airway compromise, and surgical procedure complications.

Baxter has received three complaints related to this issue, with no reported patient injury. Healthcare facilities should stop using the affected unit and contact Baxter for further instructions.

The recalled product

Product
TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087
Affected units
1

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • Serial numbers: 107971925
  • UDI-DI number: 00887761974067

Distribution

Distributed nationwide across the United States.