The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0351-2027·Announced 2026-11-11

Baxter TS7500 MOBIUS Operating Table Column Recalled for Lift Mechanism Failure

Baxter is recalling 77 hybrid operating table columns because the main lift mechanism may unexpectedly drop, posing a risk of patient injury during surgery.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, but the source describes a hazard where an uncontrolled drop of the surgical table may result in serious or critical adverse health consequences, including traumatic injury and airway compromise. The source reports three complaints but states that no patient injury was reported.

Plain-English summary

Baxter Healthcare Corporation is recalling 77 units of the TS7500 MOBIUS hybrid operating table column, Model/Catalog Number 1704695. The recall involves a defect in the main lift mechanism, which may unexpectedly drop from its set position to its lowest position during use. In such cases, the tabletop could not be raised again after the drop occurred. The issue is associated with unexpected wear of the main lift assembly, also known as the telescopic spindle.

An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. This could lead to serious or critical adverse health consequences, including traumatic injury, airway compromise, and surgical procedure complications. Baxter has received three complaints related to this issue, with no reported patient injury.

The affected units were distributed worldwide and nationwide in the United States. Healthcare facilities should identify affected units by checking the serial numbers listed in the recall notice or the UDI-DI number 00887761974425. Facilities should stop using the affected columns and contact Baxter for further instructions on repair or replacement.

The recalled product

Product
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
Affected units
77

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Serial numbers: 108094281
  • 108468075
  • 108484417
  • 108506318
  • 108516636
  • 108536785
  • 108544115
  • 108604154
  • 108463223
  • 108484404
  • 108489634
  • 108511411
  • 108516637
  • 108544106
  • 108544116
  • 108618964
  • 108463224
  • 108484405
  • 108489648
  • 108511412

Distribution

Distributed nationwide across the United States.

Similar recalls