The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0350-2027·Announced 2026-11-11

Baxter TruSystem 7500 Surgical Table Recall for Lift Mechanism Failure

Baxter is recalling 129 TruSystem 7500 surgical tables because the main lift mechanism may unexpectedly drop, potentially causing patient displacement during surgery.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, and the source describes a hazard where an uncontrolled drop may result in serious or critical adverse health consequences, including traumatic injury and airway compromise. The source reports three complaints but states there were no reported patient injuries.

Plain-English summary

Baxter Healthcare Corporation is recalling 129 units of the TruSystem 7500 surgical table, Model/Catalog Number 4091000. The recall is due to an issue with the main lift mechanism, which may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle.

An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications. Baxter has received three complaints related to this issue, with no reported patient injury.

The affected units were distributed worldwide and nationwide in the US. Users should check their serial numbers against the list provided in the official recall notice to determine if their unit is affected.

The recalled product

Product
TruSystem 7500 surgical table, Model/Catalog Number: 4091000
Affected units
129

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Serial numbers: 108491706
  • 108726496
  • 108758657
  • 108858029
  • 108945825
  • 109083212
  • 109343910
  • 109684319
  • 110016886
  • 108557062
  • 108736192
  • 108773032
  • 108882327
  • 108976094
  • 109174333
  • 109370104
  • 109690845
  • 110016887
  • 108574826
  • 108736193

Distribution

Distributed nationwide across the United States.

Similar recalls