Baxter TruSystem 7000 U Surgical Tables Recalled for Lift Mechanism Failure
Baxter is recalling 16 TruSystem 7000 U surgical tables because the main lift mechanism may unexpectedly drop, posing a risk of patient injury during surgery.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as Class II, and the source describes a hazard where the surgical table may drop, potentially causing serious adverse health consequences such as traumatic injury or airway compromise. The source reports three complaints but states there were no reported patient injuries.
Plain-English summary
Baxter Healthcare Corporation is recalling 16 units of the Operating Table TruSystem 7000 U, a universal surgical table (Model/Catalog Number 1604788). The recall is due to a defect in the main lift mechanism, which may unexpectedly drop from its set position to its lowest position during use. In such cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle.
An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications. Baxter has received three complaints related to this issue, with no reported patient injury.
The affected units were distributed worldwide and nationwide in the US. Healthcare facilities should identify affected units by checking the serial numbers listed in the recall notice or the UDI-DI number 00887761967526. Users should stop using the affected tables and contact Baxter for instructions on repair or replacement.
The recalled product
- Product
- Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788
- Manufacturer
- Baxter Healthcare Corporation
- Category
- Medical Device — Surgical Table
- Affected units
- 16
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (17)
- Serial numbers: 108517025
- 110115554
- 110144247
- 110188916
- 109558596
- 110139748
- 110166201
- 110188918
- 109737582
- 110139769
- 110177678
- 110192686
- 110115551
- 110144245
- 110185616
- 109889060
- UDI-DI number: 00887761967526
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Baxter Healthcare Corporation
See all →- SevereBaxter TruSystem 7000 U Operating Tables Recalled for Lift Mechanism Failure
FDA (Devices) · 2026-11-11
- SevereBaxter TruSystem 7000 U Operating Table Recalled for Uncontrolled Drop Risk
FDA (Devices) · 2026-11-11
- SevereBaxter TruSystem 7500 Surgical Table Recall for Lift Mechanism Failure
FDA (Devices) · 2026-11-11
- SevereBaxter TruSystem 7500 U Mobile Column Recalled for Unexpected Drop
FDA (Devices) · 2026-11-11
- SevereBaxter TS7500 MOBIUS Operating Table Column Recalled for Lift Mechanism Failure
FDA (Devices) · 2026-11-11
Same hazard · mechanical failure
See all →- ModerateCorin Unity Knee ASC Modular Handle Recall for Locking Mechanism Failure
FDA (Devices) · 2026-11-11
- Severerestor3d Recalls Veritas Reverse Total Shoulder Replacement Components
FDA (Devices) · 2026-11-11
- SevereBaxter Recalls TruSystem 7500 Operating Table Column Due to Lift Failure
FDA (Devices) · 2026-11-11
- ModerateThink Surgical Square Driver Recalled for Design Specification Failures
FDA (Devices) · 2026-10-28
Similar recalls
- SevereBaxter TruSystem 7000 U Operating Tables Recalled for Lift Mechanism Failure
FDA (Devices) · 2026-11-11
- SevereBaxter TruSystem 7000 U Operating Table Recalled for Uncontrolled Drop Risk
FDA (Devices) · 2026-11-11
- SevereBaxter TruSystem 7500 Surgical Table Recall for Lift Mechanism Failure
FDA (Devices) · 2026-11-11
- SevereBaxter Recalls TruSystem 7500 Operating Table Column Due to Lift Failure
FDA (Devices) · 2026-11-11
- SevereBaxter TruSystem 7500 U Mobile Column Recalled for Unexpected Drop
FDA (Devices) · 2026-11-11
- SevereBaxter TS7500 MOBIUS Operating Table Column Recalled for Lift Mechanism Failure
FDA (Devices) · 2026-11-11