The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0353-2027·Announced 2026-11-11

Baxter TruSystem 7000 U Operating Tables Recalled for Lift Mechanism Failure

Baxter is recalling 62 TruSystem 7000 U operating tables because the main lift mechanism may unexpectedly drop, posing a risk of patient injury during surgery.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, and the source describes a hazard where an uncontrolled drop of the surgical table may result in serious or critical adverse health consequences, including traumatic injury and airway compromise. The source reports three complaints but states that no patient injury was reported.

Plain-English summary

Baxter Healthcare Corporation is recalling 62 units of the Operating Table TruSystem 7000 U (dV), Model/Catalog Number 1723633. The recall is due to an issue with the main lift mechanism, which may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle.

An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications. Baxter has received three complaints related to this issue, with no reported patient injury.

The affected units were distributed worldwide and nationwide in the US. Users should identify affected units by checking the serial numbers listed in the recall notice or the UDI-DI number 00887761967519. Users should contact Baxter Healthcare Corporation for further instructions on how to address the issue.

The recalled product

Product
Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
Affected units
62

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Serial numbers: 108920104
  • 110192607
  • 110192885
  • 110208936
  • 110212885
  • 110220066
  • 110227520
  • 110248337
  • 110250614
  • 109423464
  • 110192608
  • 110192886
  • 110208941
  • 110212886
  • 110220068
  • 110227523
  • 110248343
  • 110250621
  • 109517138
  • 110192609

Distribution

Distributed nationwide across the United States.

Similar recalls