The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0348-2027·Announced 2026-11-11

restor3d Recalls Veritas Reverse Total Shoulder Replacement Components

restor3d Inc. is recalling specific Veritas Reverse Total Shoulder Replacement components due to a risk of dislocation or revision surgery caused by excessive mechanical loading.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II, and the source describes a mechanical defect that can result in dislocation or revision surgery. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

restor3d Inc. is recalling 240 units of Veritas Reverse Total Shoulder Replacement components, specifically the 40mm Polymer Liner paired with stem sizes 1 and 2. The affected model/catalog numbers are 3500-140C3, 3500-140U3, 3500-140C1, and 3500-140U1.

The recall was issued because pairing these specific stem sizes with the largest glenosphere diameter creates an excessive mechanical loading configuration. This condition can cause certain polyethylene components to lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or the need for revision surgery.

The products were distributed nationwide in Texas, Massachusetts, Oklahoma, Florida, Colorado, and North Carolina. The source text does not provide specific instructions for consumers or patients, but the recall is noted as retroactively reported.

The recalled product

Product
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Manufacturer
restor3d Inc.
Affected units
240

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (4)

  • All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492
  • Product Code: 3500-140U3 UDI/GTIN:00840097513461
  • Product Code: 3500-140C1 UDI/GTIN:00840097513430
  • Product Code: 3500-140U1 UDI/GTIN:00840097513409

Distribution

Distributed nationwide across the United States.

The notice also names 6 states specifically: