Radiometer ABL Software Recall for HL7 Data Transmission Errors
Radiometer America Inc. is recalling ABL software version 6.12 for ABL800 and ABL700 analyzers because HL7 2.5 data transmission errors can mix up queued results.
- Product
- The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the…
- Category
- Medical Device
- Distribution
- Distributed nationwide