Philips Brilliance 10 Slice CT Scanner Recalled for Gantry Cover Failure Risk
Philips is recalling Brilliance 10 Slice CT scanners because fiberglass resin may separate, causing the gantry front cover and metal support to fail or fall.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (falling equipment) where injury has not yet been reported, which aligns with the 'High' severity criteria in the rubric.
Plain-English summary
Philips Medical Systems (Cleveland) Inc is recalling Brilliance 10 Slice (Air) Computed Tomography (CT) X-ray systems, Model 728251. The recall is due to a potential for separation of fiberglass resin, which may cause the gantry front cover and the metal support on the cover to fail or fall.
The affected devices are distributed nationwide in the United States. The recall applies to specific system serial numbers listed in the official FDA notice. These devices are intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data.
Healthcare facilities should contact Philips Medical Systems for instructions on how to address the defect. The source text does not report any injuries or illnesses associated with this recall.
The recalled product
- Product
- Brilliance 10 Slice (Air), Model 728251 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal…
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by Philips Medical Systems (Cleveland) Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · Philips Medical Systems (Cleveland) Inc
See all →- HighRadiation planning software recalled over image errors in three patient orientations
FDA (Devices) · 2025-09-24
- HighPinnacle 3 radiation planning software may miscalculate stopping power ratio
FDA (Devices) · 2025-08-13
- HighPinnacle3 Radiation Therapy Planning System dose-calculation software error
FDA (Devices) · 2022-09-21
- SeverePhilips Forte Gamma Camera System Recalled Due to Detector Falling Hazard
FDA (Devices) · 2019-11-13
- SeverePhilips Recalls Forte Gamma Camera Systems Due to Detector Fall Risk
FDA (Devices) · 2019-11-13
Same hazard · falling object
See all →- SevereLithonia Lighting Recalls RSX Luminaires with Adjustable Tilt Arms
CPSC · 2026-10-22
- HighMaquet PowerLED 300 Concealed Lighting System Recalled for Detachment Risk
FDA (Devices) · 2026-09-30
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- SevereTerex Digger Derrick Trucks Recalled for Weld Defect
NHTSA · 2026-08-17
Similar recalls
- HighPhilips Brilliance 16 Slice CT System Recalled for Emergency Stop Defect
FDA (Devices) · 2018-04-25
- ModeratePhilips Brilliance 16 CT Scanner Recalled for Image Reconstruction Defect
FDA (Devices) · 2016-08-31
- HighPhilips CT Scanner Recall Due to Potential Cover Cracking Hazard
FDA (Devices) · 2016-07-06
- ModeratePhilips Brilliance 64 CT Scanner Recalled for Image Reconstruction Errors
FDA (Devices) · 2016-08-31
- HighPhilips Brilliance 6 Slice CT System Recalled for Emergency Stop Defect
FDA (Devices) · 2018-04-25
- ModeratePhilips Brilliance BigBore CT Scanner Recalled for Image Reconstruction Error
FDA (Devices) · 2016-08-31