Mirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
Mirion Technologies is recalling 783 Captus 4000e Thyroid Uptake Systems because a tension bolt may fail, causing the spring arm to fall unexpectedly.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a mechanical defect in a medical device that poses a risk of harm (falling arm) but does not meet the criteria for Severe (Class II with hospitalization) or Critical (Class I/fatalities).
Plain-English summary
Mirion Technologies (Capintec), Inc. is recalling 783 units of the Captus 4000e Thyroid Uptake System, including Models 5430-30151, 5430-30152, and 5430-30154. The affected devices were shipped from April 2019 through February 2024 and have serial numbers ranging from 940548 through 941398. This recall applies to devices that have not already replaced the original arm with the redesigned arm.
The recall is due to a potential failure of the tension bolt, which could allow the spring arm to unexpectedly fall to its lowest position, approximately 25 inches (63.5 cm) from the ground. The devices were distributed worldwide, including in the United States and numerous international locations.
Healthcare facilities and users should contact Mirion Technologies for instructions on how to address the defect. The FDA has classified this as a Class II recall.
The recalled product
- Product
- Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;
- Manufacturer
- Mirion Technologies (Capintec), Inc.
- Affected units
- 783
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (9)
- Affected Units: Captus 4000: 1. Model 5430-30151
- UDI-DI: 00859942006096
- 2. Model 5430-30152
- UDI-DI: 00859942006102
- 3. Model 5430-30154
- UDI-DI: 00859942006423
- Devices shipped from April 2019 through February 2024
- Serial Numbers 940548 through 941398
- that have NOT already replaced the original arm with the redesigned arm.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Mirion Technologies (Capintec), Inc.
See all →- HighMirion Captus 4000e Thyroid Uptake System Recalled for Collimator Detachment
FDA (Devices) · 2024-06-05
- HighFDA Recalls Mirion Captus 4000e Thyroid Uptake System for Collimator Detachment
FDA (Devices) · 2024-06-05
- HighCaptus 4000e Thyroid Uptake System recalled for unexpected collimator detachment
FDA (Devices) · 2024-06-05
- HighCaptus 4000e Thyroid Uptake Systems spring failure poses fall risk
FDA (Devices) · 2024-02-21
Same hazard · falling object
See all →- SevereLithonia Lighting Recalls RSX Luminaires with Adjustable Tilt Arms
CPSC · 2026-10-22
- HighMaquet PowerLED 300 Concealed Lighting System Recalled for Detachment Risk
FDA (Devices) · 2026-09-30
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- SevereTerex Digger Derrick Trucks Recalled for Weld Defect
NHTSA · 2026-08-17
- SevereIsuzu NPR HD Liftgate Deck Pin Recall
NHTSA · 2026-08-06
Similar recalls
- HighCaptus 4000e Thyroid Uptake Systems spring failure poses fall risk
FDA (Devices) · 2024-02-21
- HighCapintec Recalls Captus 3000 Thyroid Uptake Systems Due to Spring Arm Failure
FDA (Devices) · 2018-12-12
- HighCapintec Recalls Captus 700 Thyroid Uptake Systems Due to Spring Arm Failure
FDA (Devices) · 2018-12-12
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Collimator Detachment
FDA (Devices) · 2024-06-05
- HighCaptus 4000e Thyroid Uptake System recalled for unexpected collimator detachment
FDA (Devices) · 2024-06-05
- HighFDA Recalls Mirion Captus 4000e Thyroid Uptake System for Collimator Detachment
FDA (Devices) · 2024-06-05