The Recall Desk

Manufacturer

Mirion Technologies (Capintec), Inc.

4 recalls in our database name Mirion Technologies (Capintec), Inc. as the manufacturing or recalling firm.

What the numbers show

Total
4
Critical
0
Severe
0
Most recent
2024-06-05

Of the 4 recalls from Mirion Technologies (Capintec), Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–4 of 4

  • HighFDA (Devices)·Z-1894-2024·2024-06-05

    Mirion Captus 4000e Thyroid Uptake System Recalled for Collimator Detachment

    Mirion Technologies is recalling the Captus 4000e Thyroid Uptake System due to complaints of unexpected collimator detachment. Affected users should contact the manufacturer for guidance.

    Product
    Captus 4000e Thyroid Uptake System with Well and 2 inch shielding. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1892-2024·2024-06-05

    FDA Recalls Mirion Captus 4000e Thyroid Uptake System for Collimator Detachment

    The FDA is recalling the Mirion Captus 4000e Thyroid Uptake System due to reports of unexpected collimator detachment that could affect diagnostic accuracy. Approximately 1,275 units are affected worldwide.

    Product
    Captus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1893-2024·2024-06-05

    Captus 4000e Thyroid Uptake System recalled for unexpected collimator detachment

    Mirion Technologies is recalling 1,275 units of the Captus 4000e Thyroid Uptake System worldwide due to complaints of unexpected collimator detachment.

    Product
    Captus 4000e Thyroid Uptake System with Well. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipe tests.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1069-2024·2024-02-21

    Captus 4000e Thyroid Uptake Systems spring failure poses fall risk

    Mirion Technologies recalls 467 Captus 4000e Thyroid Uptake Systems due to spring arm failure that can cause the collimator to fall, posing potential injury risk to patients and operators.

    Product
    Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures. Models: 5430-30151, 5430-30152, and 5430-30154
    Category
    Medical Device
    Distribution