The Recall Desk
HighFDA (Devices)·Z-3224-2026·Announced 2026-09-30

Maquet PowerLED 300 Concealed Lighting System Recalled for Detachment Risk

Maquet is recalling 667 PowerLED 300 Concealed Lighting Systems because the spring arm may detach and fall, posing a potential injury risk.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product where the device may fall and cause injury, but no injuries or illnesses have been reported in the source text.

Plain-English summary

Maquet SAS is recalling 667 units of the PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm). The affected units are identified by Model Numbers ARD568474114A and ARD567501286. These devices were distributed nationwide in the United States.

The recall is being issued because the device has the potential to detach at the junction between the spring arm and the suspension arm. If this occurs, the device may fall, which could cause potential injury to patients or healthcare personnel.

Healthcare facilities should check their inventory for the specific model and serial numbers listed in the official recall notice. If a unit is found, it should be removed from service and returned to Maquet SAS for evaluation or replacement.

The recalled product

Product
PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;
Manufacturer
Maquet SAS
Affected units
667

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • 1. Model Number: ARD568474114A
  • UDI-DI: N/A
  • Serial Numbers: 500190
  • 500194
  • 500201
  • 500203
  • 500209
  • 500216
  • 500218
  • 500197
  • 500204
  • 500210
  • 500211
  • 500212
  • 500213
  • 500214
  • 500217
  • 500206
  • 500390
  • 500393

Distribution

Distributed nationwide across the United States.

Same manufacturer · Maquet SAS