Pinnacle3 Radiation Therapy Planning System dose-calculation software error
Philips Pinnacle3 radiation therapy planning systems may export incorrect dose information when multiple beams are used with certain Edit Prescription settings. 169 systems are affected.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall affecting a high-risk medical device (radiation therapy planning system) with a software bug that causes incorrect dose calculations. No injuries or illnesses have been reported, meeting the rubric criterion for 'risk-of-harm products where injury has not yet been reported'.
Plain-English summary
Philips Medical Systems (Cleveland) Inc is recalling 169 Pinnacle3 Radiation Therapy Planning System units (Models 870231 and 870237). The recall is due to a software error affecting dose calculation and export.
When computing a radiation dose in the system with multiple beams attached to a prescription and certain options selected on the Edit Prescription screen, the exported dose information is incorrect. This applies to Model 870231 (software versions 16.0.4, 16.4.2, and 16.4.3) and Model 870237 (software versions 16.4.6 and 18.0.1).
Affected systems were distributed across 20 US states (CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, WA) and 20 countries including Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Poland, Portugal, Romania, Russian Federation, Singapore, Spain, Switzerland, Taiwan, Thailand, and United Kingdom.
The recalled product
- Product
- Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Hazard
- dose-miscalculation
- software-error
- data-export-error
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (13)
- Model 870231: Software version 16.0.4
- UDI (01)00884838099081(10)16.0.4.60330
- Software version 16.4.2
- UDI (01)0084838103054(10)16.4.2.60330
- UDI (01)00884838102965(10)16.4.2.60330
- UDI (01)00884838102972(10)16.4.2.60315
- UDI (01)00884838102989(10)16.4.2.60330
- UDI (01)00884838103191(10)16.4.2.60330
- and Software version 16.4.3
- UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6
- UDI (01)00884838103566(10)16.4.6.21197
- and Software version 18.0.1
- UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197.
Distribution
Distributed in 21 states:
- CA
- FL
- GA
- IA
- IL
- MA
- MD
- ME
- MI
- MN
- NJ
- NM
- NV
- NY
- OH
- OK
- RI
- SC
- TN
- TX
- WA
Related recalls
Same category
- SevereMedline angiography and radiology convenience kits recalled for unapproved design changes
FDA (Devices) · 2026-07-01
- SevereDa Vinci surgical system operative components with E-brake retainer pin defect
FDA (Devices) · 2026-07-01
- ModerateOctopus 4 Tissue Stabilizer incorrectly assembled tubing positions
FDA (Devices) · 2026-07-01
- ModerateOctopus Evolution AS Tissue Stabilizer Model TS2500 Tubing Assembly Issue Recall
FDA (Devices) · 2026-07-01
- SevereMedical Device Surgical Kits with Syringes Recalled for Unapproved Design Changes
FDA (Devices) · 2026-07-01