Monaco RTP System software defect may cause incorrect radiation therapy dose calculations
Elekta's Monaco RTP System software (builds 5.40.00, 5.40.01, 5.40.02, 5.51.10) may calculate incorrect radiation doses if a treatment couch is removed and reinserted during specific plan loading operations.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This medical device recall involves a software defect in a radiation therapy planning system that could result in incorrect dose calculations. No reports of actual patient harm are mentioned in the recall notice, classifying this as a risk-of-harm product without reported injury, resulting in a High (3) severity per the established rubric.
Plain-English summary
Elekta Inc. is recalling specific software builds of the Monaco RTP System, a radiation therapy planning system used to calculate treatment doses for cancer patients. The affected software builds are 5.40.00, 5.40.01, 5.40.02, and 5.51.10.
A software defect has been identified in these builds. If a treatment couch (patient support) is removed while a treatment plan is loaded into the system, and the couch is then added back when the plan is unloaded, the system may calculate radiation doses with the couch improperly assigned to the radiation beams. This could result in inaccurate dose calculations.
The Monaco RTP System has been distributed in the United States in Iowa, New Jersey, New York, Pennsylvania, Tennessee, Texas, and Wisconsin. International distribution includes Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey, and the United Kingdom.
Facilities using affected builds of the Monaco RTP System should contact Elekta Inc. for instructions on obtaining a software update that corrects this defect. Users should review affected treatment plans to ensure accurate dose calculations.
The recalled product
- Product
- Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10
- Manufacturer
- Elekta Inc
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Software Builds: 5.40.00
- 5.40.01
- 5.40.02
Distribution
Related recalls
Same manufacturer · Elekta Inc
See all →- SevereBrachytherapy Applicator Advanced Gynecological Applicator Tip Detachment
FDA (Devices) · 2026-07-15
- SevereElekta Geneva Brachytherapy Applicator Intrauterine Tube Detachment Risk
FDA (Devices) · 2026-07-15
- SevereFletcher CT/MR Shielded Applicator Set for Gynecological Brachytherapy Recalled
FDA (Devices) · 2026-07-15
- HighMOSAIQ oncology software may fail to load adapted treatment plan
FDA (Devices) · 2026-06-17
- HighLeksell Vantage Arc System locking mechanism may not secure properly
FDA (Devices) · 2026-05-06
Same hazard · dose miscalculation
See all →- HighRayStation radiation therapy software recall for density calculation inconsistency
FDA (Devices) · 2025-02-26
- HighPinnacle3 Radiation Therapy Planning System dose-calculation software error
FDA (Devices) · 2022-09-21
- HighRadiotherapy Planning Software BeamAdjust and VeriSoft Recalled for Measurement Errors
FDA (Devices) · 2022-01-19
- SevereRaySearch Ray Station Software Recalled for Dose Miscalculations
FDA (Devices) · 2014-04-30
Similar recalls
- HighMonaco RTP System software may calculate radiation therapy doses inaccurately
FDA (Devices) · 2023-03-15
- HighElekta Monaco RTP System Recalled for Potential Dose Calculation Errors
FDA (Devices) · 2019-10-02
- SevereElekta Monaco RTP System Recalled for Incorrect Dose Calculation Errors
FDA (Devices) · 2017-01-25
- HighElekta Monaco Radiation Treatment Planning System Software Recall
FDA (Devices) · 2019-12-18
- SevereElekta MONACO RTP System Recalled for Potential Inaccurate Radiation Dose Delivery
FDA (Devices) · 2020-05-06
- HighElekta Monaco RTP System Recalled for Incorrect Dose Calculation
FDA (Devices) · 2016-10-19