  • ModerateFDA (Devices)·Z-1632-2023·2023-05-31

    Siemens Atellica CH Toxicology Calibrator Reassignment Due to Positive Bias

    Siemens Healthcare Diagnostics is reassigning Atellica CH Toxicology Calibrators due to a positive bias. The affected calibrators were distributed in the US and internationally across multiple countries.

    Product
    Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica¿ CH Analyzer Siemens Material Number (SMN): 11099440
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0767-2023·2023-05-31

    Fentanyl Buccal Tablets Recalled for Missing or Incorrect Package Insert

    Teva Pharmaceuticals is recalling Fentanyl Buccal Tablets (200mcg) due to missing or incorrect package inserts. Patients should contact their healthcare provider if they have affected product.

    Product
    Fentanyl Buccal Tablets CII, 200mcg, packaged in cartons of 28 Buccal Tablets (4 tablets x 7 cards), Rx only, Distributed by: Mayne Pharma, Greenville, NC 27834, NDC 51862-635-28
    Category
    Drug
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1480-2023·2023-05-31

    Fluoroscopic Imaging System May Display Incorrect Radiation Dose Measurements

    Siemens Luminos dRF Max fluoroscopic imaging systems may occasionally display incorrect radiation dose measurements during certain operating conditions. The manufacturer states the error does not affect diagnosis or workflow.

    Product
    Luminos dRF Max (VE10, VF10, VF11)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1481-2023·2023-05-31

    LUMINOS Lotus Max fluoroscopic system recalled for sporadic measurement errors

    LUMINOS Lotus Max fluoroscopic systems may sporadically display incorrect radiation dose measurements. The error only occurs with secondary X-ray tubes and does not affect diagnosis or workflow.

    Product
    LUMINOS Lotus Max (VF11)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0757-2023·2023-05-31

    Nationwide Recall of Glimepiride Tablets for Manufacturing Practice Deviations

    Amerisource Health Services LLC recalls 63,335 bottles of Glimepiride Tablets nationwide due to manufacturing practice deviations identified during FDA inspection.

    Product
    Glimepiride Tablets, USP, 1 mg, RX, Packaged as a)100-count bottle, NDC# 68001-177-00; b) 500-count bottle, NDC# 68001-177-03; Manufactured by: Intas Pharmaceuticals Limited, Pharmez, Ahmedabad-382 213. INDIA. For BluePoint Laboratories
    Category
    Drug
    Distribution
    Distributed nationwide
  • LowFDA (Drugs)·D-0770-2023·2023-05-31

    Fentanyl buccal tablets recalled due to incorrect or missing package insert

    Teva Pharmaceuticals USA Inc is recalling Fentanyl Buccal Tablets 800mcg due to incorrect or missing package inserts. The recall affects 3,032 units distributed nationwide.

    Product
    Fentanyl Buccal Tablets CII, 800mcg, packaged in cartons of 28 Buccal Tablets (4 tablets x 7 cards), Rx only, Distributed by: Mayne Pharma, Greenville, NC 27834, NDC 51862-638-28
    Category
    Drug
    Distribution
    Distributed nationwide
  • LowFDA (Drugs)·D-0766-2023·2023-05-31

    Fentanyl Buccal Tablets recalled due to missing or incorrect package insert

    Teva Pharmaceuticals is recalling Fentanyl Buccal Tablets 100mcg due to incorrect or missing package inserts in affected cartons. Patients should verify proper labeling or contact their pharmacy.

    Product
    Fentanyl Buccal Tablets CII, 100mcg, packaged in cartons of 28 Buccal Tablets (4 tablets x 7 cards), Rx only, Distributed by: Mayne Pharma, Greenville, NC 27834, NDC 51862-634-28
    Category
    Drug
    Distribution
    Distributed nationwide
  • LowFDA (Drugs)·D-0769-2023·2023-05-31

    Fentanyl Buccal Tablets 600mcg Recalled for Incorrect or Missing Package Insert

    Teva Pharmaceuticals is recalling Fentanyl Buccal Tablets 600mcg due to incorrect or missing package inserts in 1,908 units distributed nationwide across multiple lot numbers.

    Product
    Fentanyl Buccal Tablets CII, 600mcg, packaged in cartons of 28 Buccal Tablets (4 tablets x 7 cards), Rx only, Distributed by: Mayne Pharma, Greenville, NC 27834, NDC 51862-637-28
    Category
    Drug
    Distribution
    Distributed nationwide
  • LowFDA (Drugs)·D-0756-2023·2023-05-31

    Methylprednisolone Acetate Injectable Suspension recalled for NDC carton labeling error

    Amneal Pharmaceuticals recalls Methylprednisolone Acetate Injectable Suspension due to incorrect NDC number on product cartons. The typographical error appears on secondary packaging only; the medication vial itself is unaffected.

    Product
    Methylprednisolone Acetate Injectable Suspension, USP 80 mg/mL, For Intramuscular, Intrasynovial and Soft Tissue Injection Only, Not for Intravenous Use, 1 mL Single Dose Vial, Manufactured by: Amneal Pharmaceuticals Pvt. Ltd. Parenteral Unit, Ahmedabad 382213, India, Amneal…
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereNHTSA·23V382000·2023-05-30

    Alfa Romeo Giulia and Stelvio Recalled for Brake Rotor Fracture Risk

    NHTSA is recalling 2017-2020 Alfa Romeo Giulia and 2018-2020 Stelvio vehicles with carbon ceramic brakes due to a risk of rotor fracture and decreased brake performance.

    Product
    ALFA ROMEO — ALFA ROMEO GIULIA, STELVIO
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereNHTSA·23V384000·2023-05-30

    2023 Toyota Corolla Cross instrument panel airbag deployment risk

    Toyota is recalling 2022-2023 Corolla Cross vehicles because the instrument panel may lack required perforations, preventing proper front-passenger airbag deployment in a crash.

    Product
    TOYOTA — TOYOTA COROLLA CROSS
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereNHTSA·23V378000·2023-05-26

    Ford Recalls 2015-2019 Lincoln MKC for Engine Compartment Fire Risk

    Ford is recalling 2015-2019 Lincoln MKC vehicles because a battery monitor sensor may short-circuit and overheat, causing a fire in the engine compartment.

    Product
    LINCOLN — LINCOLN MKC
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V375000·2023-05-26

    Forest River Recalls 2023 East to West Travel Trailers Due to Gas Leak Risk

    Forest River is recalling 2023 East to West Della Terra and Silver Lake travel trailers because an incorrectly installed furnace may cause a gas leak inside the cabin.

    Product
    EAST TO WEST — EAST TO WEST DELLA TERRA, SILVER LAKE
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V381000·2023-05-26

    Knapheide Recalls Vehicles with Warn Winches Due to Unintentional Operation Risk

    Knapheide is recalling 2022-2023 trucks equipped with certain Warn winches because water intrusion can cause the winch to operate unintentionally, increasing injury risk.

    Product
    KNAPHEIDE — KNAPHEIDE RAM 3500, FORD F-350, FORD F-250, FORD F-550
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V380000·2023-05-26

    Ford Escape, Maverick, and Corsair Engine Fire Risk Recall

    Ford is recalling certain 2020-2023 Escape, 2022-2023 Maverick, and 2021-2023 Corsair vehicles due to a 2.5L engine defect that could cause an engine compartment fire. No injuries have been reported.

    Product
    FORD — FORD, LINCOLN MAVERICK, CORSAIR, ESCAPE
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V379000·2023-05-26

    Ford Vehicles Recalled for Incorrect Tire and Loading Capacity Label

    Ford is recalling certain 2019-2023 Ranger, 2022-2023 Bronco, and 2023 F-250 SD vehicles with TREAD labels stating incorrect load carrying capacity, which could result in vehicle overload and increase crash risk.

    Product
    FORD — FORD F-250 SD, RANGER, BRONCO
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V376000·2023-05-26

    Travel trailers recalled for missing wardrobe cabinet backer board

    Certain 2021-2023 Keystone Arcadia travel trailers lack a backer board behind the wardrobe cabinet, potentially allowing the slide room to extend and cause injury. Dealers will install steel brackets to secure the slide room at no cost.

    Product
    KEYSTONE — KEYSTONE ARCADIA
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereNHTSA·23V373000·2023-05-25

    2021-2023 Autocar Xpeditor trucks recalled for defective door latch

    Autocar is recalling 2021-2023 Xpeditor commercial trucks due to incorrectly assembled right-side door latches that may open unexpectedly during operation. Owners should contact Autocar for a free inspection and reassembly.

    Product
    AUTOCAR — AUTOCAR XPEDITOR
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V369000·2023-05-25

    Jaguar I-PACE Battery Overheat May Increase Fire Risk

    Jaguar is recalling certain 2019-2024 I-PACE vehicles because the high-voltage battery may overheat, increasing the risk of fire. Owners should park and charge vehicles outside until repairs are complete.

    Product
    JAGUAR — JAGUAR I-PACE
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V368000·2023-05-25

    Fontaine Recalls 2023-2024 Container Trailers Due to Missing Reflectors

    Fontaine Trailer Company is recalling 2023-2024 40-foot container trailers because they may be missing rear reflex reflectors, which reduces visibility and increases crash risk.

    Product
    FONTAINE — FONTAINE CONTAINER TRAILER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V371000·2023-05-25

    IC Bus school buses recalled for ABS warning light failure

    IC Bus school buses (2021–2024) are being recalled because the anti-lock brake system warning light fails to illuminate during malfunctions, preventing drivers from being alerted and increasing crash risk.

    Product
    IC BUS — IC BUS CESB, RESB, EVSB
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V367000·2023-05-25

    2023 Cadillac Lyriq Battery Connection Defect May Cause Loss of Drive Power

    General Motors is recalling certain 2023 Cadillac Lyriq and 2022 GMC Hummer EV vehicles. Connections in the high-voltage battery pack may be out of position or incorrectly welded, potentially causing loss of drive power and increasing crash risk.

    Product
    CADILLAC — CADILLAC, GMC LYRIQ, HUMMER EV PICKUP
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighCPSC·23210·2023-05-25

    Janod Sweet Cocoon Activity Tables Recalled for Choking Hazard

    Juratoys is recalling about 5,080 Janod Sweet Cocoon Activity Tables because a detachable silver metal bell poses a choking hazard to children 12 months and older. Consumers should stop using the tables and contact Juratoys for a free repair kit.

    Product
    Janod Sweet Cocoon Activity Tables
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • HighNHTSA·23V374000·2023-05-25

    2023 Nissan Rogue rear seat belt fasteners may loosen in crashes

    Certain 2023 Nissan Rogue vehicles have rear seat belt anchor fasteners that may not have been properly tightened and could loosen over time, potentially failing to restrain occupants in a crash.

    Product
    NISSAN — NISSAN ROGUE
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·23V370000·2023-05-25

    2022-2024 International Trucks: ABS Warning Light Fails to Illuminate

    Navistar is recalling certain 2021-2024 International trucks and buses whose anti-lock brake system warning light fails to illuminate when malfunctions occur. Drivers will not be alerted to brake system problems, increasing the risk of crashes.

    Product
    INTERNATIONAL — INTERNATIONAL, IC BUS LONESTAR, RH, LT, HX
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighCPSC·23761·2023-05-25

    Bunch Bikes Recalls The Preschool Electric Bicycles Due to Excess Lead Paint

    Bunch Bikes is recalling about 150 units of The Preschool Electric Bicycles sold online from April 2020 through December 2022 because the paint contains lead levels exceeding the federal ban. Lead is toxic if ingested by young children and can cause adverse health effects.

    Product
    The Preschool Electric Bicycles
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • ModerateCPSC·23762·2023-05-25

    The Good and the Beautiful Math Boxes Recalled Due to Laceration Hazard

    The Good and the Beautiful is recalling Math 1 and Math 3 boxes with metallic whiteboards because the attached whiteboards can detach and cause lacerations. Four minor laceration injuries have been reported.

    Product
    The Good and the Beautiful Math 1 and Math 3 Boxes with metallic whiteboards
    Category
    Consumer Product
    Distribution
    Distributed nationwide
  • CriticalFDA (Food)·F-0920-2023·2023-05-24

    Gold Medal Flour Recalled Nationwide for Potential Salmonella Contamination

    General Mills is recalling Gold Medal Bleached All Purpose Flour nationwide due to potential Salmonella contamination. Approximately 271,454 bags are affected.

    Product
    Gold Medal Bleached All Purpose Flour, Enriched, Presifted, packaged in the following sizes: 1. Net Wt 32 oz (2lb) 907g, UPC 0 16000-10710 6; and 2. Net Wt 5lb (2.26kg), UPC 0 16000-10610 9. Dist. By General Mills Sales, Inc., Minneapolis, MN 55440.
    Category
    Food
    Distribution
    Distributed nationwide
  • CriticalFDA (Food)·F-0885-2023·2023-05-24

    Highkey Soft Baked Mini Treats recalled for undeclared egg allergen

    Highkey Soft Baked Mini Treats Birthday Cake was recalled because egg allergen was not listed on the label, causing an allergic reaction in a consumer. The product was distributed to seven U.S. states.

    Product
    Highkey Soft Baked MINI TREATS BIRTHDAY CAKE NET WT 2 OZ (57g) CONTAINS: ALMONDS, MILD, AND COCONUT. DISTRIBUTED BY SUMMIT NATURALS INC ORLANDO, FL 32826 UPC 8 50020 43374 2
    Category
    Food
    Distribution
    7 states
  • CriticalFDA (Devices)·Z-1447-2023·2023-05-24

    Draeger VentStar Anesthesia Breathing Circuit Connections May Detach

    Draeger Medical recalls VentStar Anesthesia (N) 180 breathing circuits. Glued connections can loosen or detach during ventilation, potentially creating a critical disconnection risk.

    Product
    VentStar Anesthesia (N) 180, MP00333
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • CriticalFDA (Food)·F-0893-2023·2023-05-24

    SW Texas Nut Blend Recalled for Undeclared Almonds and Cashews

    SW Texas Nut Blend (Item #5180819) is being recalled due to undeclared almonds and cashews. The product label lists only hazelnuts and walnuts as ingredients.

    Product
    SW Texas Nut Blend, Item #5180819, Packed in cardboard box with poly liner; net wt. 25 lbs. No UPC. Label declares: SW Texas Nut Blend, ingredients: Hazelnuts, Walnuts. Manufactured by: Ardent Mills Ave. Tualatin, OR.
    Category
    Food
    Distribution
    1 state
  • SevereFDA (Devices)·Z-1590-2023·2023-05-24

    Impella 5.5 with SmartAssist Catheter Recalled for Heightened Purge Leak Rate

    Abiomed is recalling 610 units of the Impella 5.5 with SmartAssist catheter due to a heightened complaint rate for purge leaks. This intravascular pump provides circulatory support during critical care.

    Product
    Impella 5.5 with SmartAssist catheter is an intravascular micro axial blood pump that supports a patient s circulatory system, Product Numbers 1000211 (Impella 5.5 JP), 0550-0002 (Impella 5.5 Smart Assist), and 0550-0008 (Impella 5.5 with SmartAssist Set, US)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1460-2023·2023-05-24

    Anesthesia circuit kit may detach during ventilation due to loose glued connections

    Draeger Medical is recalling its Anesthesia Circuit Kit Coax HEPA (720 units) because glued connections can become loose during ventilation, potentially causing partial or complete detachment that could interrupt oxygen delivery during surgery.

    Product
    Anesthesia Circuit Kit Coax HEPA, MP17102
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereNHTSA·23E040000·2023-05-24

    Veoneer Brake Booster Recall for Certain Honda Vehicles

    Veoneer is recalling brake booster assemblies in certain Honda vehicles because a tie rod fastener may be improperly assembled, risking brake failure.

    Product
    SERVICE BRAKES recall
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1458-2023·2023-05-24

    Anesthesia Set Coax 180 Breathing Circuit Connection Risk

    Draeger Medical is recalling the Anesthesia Set Coax 180 because glued breathing circuit connections can become loose during ventilation, causing partial or complete detachment.

    Product
    Anesthesia Set Coax 180, MP03384
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1442-2023·2023-05-24

    VentStar Watertrap Breathing Circuit Connections May Detach During Use

    Draeger Medical recalls VentStar Watertrap (P) 180 breathing circuit equipment because glued connections can become loose or detach before or during ventilation.

    Product
    VentStar Watertrap (P) 180, MP00361
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Food)·F-0954-2023·2023-05-24

    Tahini Recall: Salmonella Contamination Risk in Mighty Sesame Organic Product

    Kenover Marketing Corp. is recalling Mighty Sesame Organic Tahini due to potential Salmonella contamination. The affected product was distributed to multiple states.

    Product
    Mighty Sesame 10.9 Oz Organic Tahini (Squeezable) bottle, UPC 858313006208, 8 Units per case.
    Category
    Food
    Distribution
    16 states
  • SevereFDA (Devices)·Z-1457-2023·2023-05-24

    Ventstar Coax 230 breathing circuit connections can detach during ventilation

    The FDA is recalling Draeger Medical's Ventstar Coax 230 breathing circuit components due to glued connections that can become loose and detach during ventilation, potentially compromising patient breathing support.

    Product
    Ventstar Coax 230, MP03376
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1455-2023·2023-05-24

    Ventstar Coax 180 Ventilator Breathing Circuit Connections May Detach

    Draeger Medical is recalling 18,160 Ventstar Coax 180 ventilators because glued breathing circuit connections can become loose and detach during use, potentially disrupting oxygen delivery.

    Product
    Ventstar Coax 180, MP03374
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1446-2023·2023-05-24

    ID Circuit Basic breathing circuit recalled for loose glued connections

    Draeger Medical is recalling ID Circuit Basic breathing circuits due to glued connections that can loosen during ventilation, potentially causing partial or complete detachment of circuit pieces.

    Product
    ID Circuit Basic (P) 180, MP01340
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1445-2023·2023-05-24

    VentStar Breathing Bag Set Recalled for Loose Circuit Connections

    Draeger Medical is recalling VentStar breathing bag sets nationwide due to loose glued connections in the breathing circuit. Connections can detach before or during ventilation.

    Product
    VentStar breathing bag Set (P) 110, MP00384
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Food)·F-0919-2023·2023-05-24

    Gold Medal Unbleached All Purpose Flour Recalled for Potential Salmonella Contamination

    General Mills is recalling Gold Medal Unbleached All Purpose Flour for potential Salmonella contamination. The recall affects 142,408 bags distributed nationwide with best-by dates of March 27–28, 2024.

    Product
    Gold Medal Unbleached All Purpose Flour, Enriched, Presifted, packaged in the following sizes: 1. Net Wt 5 lb (2.26kg), UPC 0 16000-19610 0; and 2. Net Wt 10 lb (4.53kg), UPC 0 16000-19580 6. Dist. By General Mills Sales, Inc., Minneapolid, MN 55440.
    Category
    Food
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1449-2023·2023-05-24

    VentStar Watertrap breathing circuit connections may detach during ventilation

    Draeger Medical is recalling the VentStar Watertrap (N) 180 due to glued breathing circuit connections that may become loose or detach during ventilation, posing patient safety risk.

    Product
    VentStar Watertrap (N) 180, MP00363
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1440-2023·2023-05-24

    Draeger VentStar Basic ventilator breathing circuit connections may detach during use

    Draeger Medical is recalling 325 VentStar Basic (P)180 ventilators because glued breathing circuit connections can become loose before or during ventilation, potentially causing partial or complete detachment of circuit pieces.

    Product
    VentStar Basic (P)180, MP00351
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1448-2023·2023-05-24

    VentStar Basic Ventilator Recall: Risk of Breathing Circuit Detachment

    Draeger Medical is recalling 1,725 VentStar Basic ventilators because glued breathing circuit connections can become loose and detach before or during ventilation. This may disrupt patient breathing support.

    Product
    VentStar Basic (N) 180, MP00353
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1566-2023·2023-05-24

    Plum A+ & A+3 Infusion System Batteries Recalled Due to Manufacturing Defect

    ICU Medical Inc is recalling replacement batteries used in Plum A+ and A+3 infusion pump systems. A manufacturing defect may cause batteries to lose capacity earlier than expected.

    Product
    Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1565-2023·2023-05-24

    Replacement Battery in Plum 360 Infusion System Recalled for Capacity Loss

    ICU Medical Inc is recalling 28,656 replacement batteries used in the Plum 360 Infusion System due to a manufacturing defect that causes premature capacity loss and decreased runtime. Affected units are distributed worldwide.

    Product
    Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1567-2023·2023-05-24

    ASM Replacement Battery Recalled for Manufacturing Defect and Capacity Loss

    ICU Medical is recalling the ASM Replacement Battery due to a manufacturing defect from the battery supplier. The defect may cause batteries to lose capacity and experience decreased runtime earlier than expected.

    Product
    ASM Replacement Battery, Component Number SUB0000864
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-1451-2023·2023-05-24

    VentStar Coax breathing circuit connections recalled for detachment risk

    Draeger Medical is recalling VentStar Coax (P) 150 breathing circuit components due to potentially loose glued connections that could detach during ventilation, affecting 280 units distributed nationwide.

    Product
    VentStar Coax (P) 150, MP00379
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1568-2023·2023-05-24

    ASM Replacement Battery Recalled for Manufacturing Defect and Capacity Loss

    ICU Medical is recalling ASM Replacement Batteries due to a manufacturing defect from the battery supplier. Affected batteries may lose capacity and runtime earlier than expected during their lifecycle.

    Product
    ASM Replacement Battery, Component Number SUB0000594
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Food)·F-0890-2023·2023-05-24

    Ground Cumin Distributed by Lipari Foods Recalled for Salmonella

    Lipari Foods is recalling 147 cases of ground cumin contaminated with Salmonella. The product was distributed to retail stores across 16 states.

    Product
    Distributed by Lipari Foods Ground Cumin in 6oz plastic tub, UPC 094776212620, 12 retail units per wholesale case
    Category
    Food
    Distribution
    16 states
  • SevereFDA (Devices)·Z-1444-2023·2023-05-24

    Anesthesia Ventilator Breathing Circuit Connections May Detach During Use

    Draeger Medical is recalling 130 units of the VentStar Anesthesia WT ventilator due to a defect where glued breathing circuit connections may loosen or detach during ventilation, potentially compromising anesthesia delivery.

    Product
    VentStar Anesthesia WT (P)180, MP00374
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1441-2023·2023-05-24

    VentStar Basic ventilator breathing circuit connections may detach

    Draeger Medical is recalling 200 units of the VentStar Basic ventilator due to glued breathing circuit connections that can loosen or detach before or during ventilation.

    Product
    VentStar Basic (P)250, MP00352
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Food)·F-0884-2023·2023-05-24

    Highkey Banana Nut Muffin Mini Treats Recalled for Undeclared Egg Allergen

    Highkey soft baked MINI TREATS BANANA NUT MUFFIN are being recalled due to undeclared egg on the label. A consumer experienced an allergic reaction to the product.

    Product
    Highkey soft baked MINI TREATS BANANA NUT MUFFIN NET WT 2 OZ (57g) CONTAINS: ALMONDS, MILK, AND PECANS. DISTRIBUTED BY SUMMIT NATURALS INC ORLANDO, FL 32826 UPC 8 50020 43373 5
    Category
    Food
    Distribution
    7 states
  • SevereFDA (Devices)·Z-1453-2023·2023-05-24

    Anesthesia Circuit Kit Coax 3 Connections May Detach During Ventilation

    Draeger Medical is recalling Anesthesia Circuit Kit Coax 3 units because glued connections can become loose or detach before or during ventilation. The recall covers 329,294 units distributed nationwide.

    Product
    Anesthesia Circuit Kit Coax 3, MP02732
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1450-2023·2023-05-24

    Draeger VentStar Breathing Circuit Connections May Loosen During Use

    Draeger Medical is recalling the VentStar Bag Set because glued connections on the breathing circuit can become loose or detach during ventilation, potentially compromising respiratory support.

    Product
    VentStar Bag Set (N) 110, MP00383
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Food)·F-0892-2023·2023-05-24

    Oregon Nut Blend Recalled for Undeclared Almonds and Cashews

    Willamette Hazelnut, Inc. is recalling Oregon Nut Blend because it contains undeclared almonds and cashews. This poses a risk to consumers allergic to these nuts.

    Product
    Oregon Nut Blend, Item # 5180808, Packed in cardboard box with poly liner; net wt. 25 lbs. No UPC. Label declares: Oregon Nut Blend, Innovative Bakery Resources. Ingredients: Hazelnuts, Walnuts. Product of USA. Manufactured By: Ardent Mills Tualatin, OR.
    Category
    Food
    Distribution
    1 state
  • HighFDA (Devices)·Z-1598-2023·2023-05-24

    Urine Microscopy Analyzer Software Defect May Prevent Carryover Detection

    Beckman Coulter is recalling the DxU 840m Iris Urine Microscopy Analyzer due to a software issue where a carryover detection flag may fail to display, potentially resulting in undetected sample contamination and incorrect test results.

    Product
    DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Drugs)·D-0641-2023·2023-05-24

    Vancomycin Injectable Antibiotic Recalled Due to Sterility Concerns

    Apollo Care, LLC is recalling VANComycin 1.75 g injectable solution (NDC 71170-194-50) due to lack of assurance of sterility. The recall affects 240 bags distributed in Missouri.

    Product
    VANComycin 1.75 g added to 500 mL of 0.9% Sodium Chloride Injection, Rx Only, Apollo Care, LLC, 3801 Mojave Ct., Suite 101, Columbia, MO 65202. NDC 71170-194-50
    Category
    Drug
    Distribution
    1 state
  • HighFDA (Devices)·Z-1600-2023·2023-05-24

    Fisher Wallace Stimulator Models Recalled for Battery Overheat and Burn Risk

    Fisher Wallace Stimulator Models FW-100 and FW-200 are being recalled because incorrectly installed batteries may cause the device to overheat, potentially resulting in burns. The recall affects 43,489 devices distributed nationwide and globally.

    Product
    Fisher Wallace Stimulator Models FW-100 and FW-200
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0602-2023·2023-05-24

    Ampicillin and Sulbactam Injection Recalled Due to Sterility Assurance Deficiency

    Ampicillin and Sulbactam for Injection is being recalled nationwide because the manufacturer cannot assure sterility of 591,160 vials distributed throughout the United States.

    Product
    Ampicillin and Sulbactam for Injection, USP 3 grams per vial, Rx only, Manufactured for: Xellia Pharmaceuticals USA, LLC, Buffalo Grove, IL 60089, NDC 70594-082-02
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0614-2023·2023-05-24

    Ceftriaxone Injectable Antibiotic Recalled for Lack of Sterility Assurance

    Astral SteriTech is recalling 41,500 vials of Ceftriaxone for Injection due to lack of assurance of sterility. The antibiotic was distributed nationwide.

    Product
    Ceftriaxone for Injection, USP 2 grams per vial, Rx only, Mfg. for: Piramal Critical Care, Bethlehem, PA 18017, NDC 66794-214-42
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0632-2023·2023-05-24

    PHENYLephrine HCl Injection Recalled for Lack of Sterility Assurance

    Apollo Care, LLC is recalling 2,265 syringes of PHENYLephrine HCl Injection (100 mcg/mL) due to lack of assurance of sterility. The affected lots were distributed in Missouri only.

    Product
    PHENYLephrine HCl Injection, 1 mg per 10 mL (100 mcg/mL), Single-Use Syringe, Solution for IV Use Only. Apollo Care, LLC, 3801 Mojae Ct., Suite 101, Columbia, MO 65202. NDC: 71170-010-10
    Category
    Drug
    Distribution
    1 state
  • HighFDA (Drugs)·D-0576-2023·2023-05-24

    Lorazepam Injection Vials Recalled Due to Manufacturing Quality Control Failures

    RemedyRepack Inc. is recalling 9,038 vials of Lorazepam Injection due to discontinuation of the manufacturer's quality control program. The product may not meet specified identity, strength, quality, and purity standards.

    Product
    LORazepam Injection, 2mg / mL Single Dose vial 1ml vial, Rx Only, MFG Akorn Lake Forest IL 60045, Repackaged by: RemedyRepack Inc., Indiana PA 15701, Source NDC # 17478-0040-01, Remedy NDC 70518-2268-00
    Category
    Drug
    Distribution
    1 state
  • HighFDA (Drugs)·D-0605-2023·2023-05-24

    Cefepime for Injection Vials Recalled Over Sterility Concerns

    Cefepime for Injection vials are being recalled nationwide due to lack of sterility assurance. The product should not be used and affected individuals should contact their healthcare provider.

    Product
    Cefepime for Injection, USP 2 grams per vial, Rx only, Manufactured for: Civica Inc. Lehi Utah 84043, NDC 72572-058-10
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1601-2023·2023-05-24

    Surgical Optical Trocars Recalled Due to Sterility Assurance Failure

    Surgical Innovations Ltd is recalling 2380 YelloPort Elite optical trocars due to inability to guarantee sterility. Patients using contaminated surgical instruments may develop serious infections.

    Product
    YelloPort Elite, REF: EO0507005, 5x70 Optical Trocar, , Rx Only, SterileR
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Drugs)·D-0622-2023·2023-05-24

    Piperacillin and Tazobactam Injectable Recalled for Sterility Assurance Issue

    Astral SteriTech is recalling 204,860 vials of Piperacillin and Tazobactam for Injection due to lack of assurance of sterility. The FDA classified this as a Class II recall affecting nationwide distribution.

    Product
    Piperacillin and Tazobactam for Injection, USP, 4.5 grams per vial, Rx only, Manufactured for: Civica Inc. Lehi Utah 84043, NDC 72572-577-10
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0582-2023·2023-05-24

    Ampicillin Injection Vials Recalled for Sterility Assurance Failure

    Astral SteriTech is recalling approximately 89,690 vials of Ampicillin for Injection due to lack of assurance of sterility. The product was distributed nationwide.

    Product
    Ampicillin for Injection, USP 1g per vial, Rx only, Mfg. for: Piramal Critical Care, Bethlehem, PA 18017 USA, NDC 66794-222-41
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1603-2023·2023-05-24

    Immundiagnostik Lactoferrin ELISA Kit Recalled for Lacking FDA 510(k) Clearance

    IMMUNDIAGNOSTIK Inc. has recalled the IDK Lactoferrin ELISA diagnostic kit because it was marketed without the required FDA 510(k) premarket notification. No illnesses or injuries have been reported.

    Product
    Immundiagnostik IDK Lactoferrin ELISA- In vitro enzyme immunoassay intended for the determination of lactoferrin in stool. P/N: K6870 Single Kit P/N: K6870.20.US.L 20 Plates
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1596-2023·2023-05-24

    Urine Microscopy Analyzer Software Display Defect May Prevent Contamination Detection

    Beckman Coulter's iQ200 Series urine analyzers may fail to display a flag indicating prior samples, potentially preventing detection of sample contamination from carryover events.

    Product
    iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Drugs)·D-0629-2023·2023-05-24

    Piperacillin and Tazobactam for Injection Recalled for Lack of Sterility Assurance

    The FDA is recalling 169,520 vials of Piperacillin and Tazobactam injectable antibiotic due to lack of assurance of sterility. The affected vials were distributed nationwide.

    Product
    Piperacillin and Tazobactam for Injection, USP 2.25 g per vial,s Rx only, Manufactured for: Xellia Pharmaceuticals USA, LLC, Buffalo Grove, IL 60089, NDC 70594-078-02
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0592-2023·2023-05-24

    Ampicillin-Sulbactam Injection Recalled Due to Sterility Concerns

    FDA recalls Ampicillin and Sulbactam for Injection (1.5 grams per vial) manufactured by Astral SteriTech due to lack of sterility assurance. The recall affects 236,610 vials distributed nationwide.

    Product
    Ampicillin and Sulbactam for Injection, USP 1.5 grams per vial, Rx only, Mfd for Meitheal Pharmaceuticals Chicago, IL 60631, NDC 71288-005-20
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0633-2023·2023-05-24

    Succinylcholine Injection recalled for lack of sterility assurance

    Apollo Care, LLC is recalling Succinylcholine Cl Injection due to lack of sterility assurance. The affected lot was distributed in Missouri.

    Product
    Succinylcholine Cl Injection, 100 mg per 5 mL (20 mg/mL), Single-Use Syringe, Rx only, Apollo Care, LLC, 3801 Mojave Ct., Suite 101, Columbia, MO 65202. NDC 71170-020-05
    Category
    Drug
    Distribution
    1 state
  • HighFDA (Devices)·Z-1580-2023·2023-05-24

    Breathing circuit swivel elbow recalled for inadequate sterilization instructions

    Smiths Medical is recalling DuraLife Swivel Elbow breathing circuit devices due to inadequate sterilization instructions. The device instructions lack cleaning methods, autoclave cycle duration, and reuse specifications.

    Product
    DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with breathing circuits
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0630-2023·2023-05-24

    Piperacillin and Tazobactam Injection Recalled Due to Sterility Assurance Issue

    Approximately 936,680 vials of Piperacillin and Tazobactam for Injection are being recalled nationwide due to lack of assurance of sterility. All affected lots remain within their expiration dates.

    Product
    Piperacillin and Tazobactam for Injection, UPS, 3.375 grams per vial, Rx only, Manufactured for: Xellia Pharmaceuticals USA, LLC, Buffalo Grove, IL 60089, NDC 70594-079-02
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0625-2023·2023-05-24

    Piperacillin and Tazobactam Injection Recalled for Sterility Assurance Failure

    Meitheal Pharmaceuticals is recalling Piperacillin and Tazobactam for Injection due to lack of sterility assurance. The recalled product may pose an infection risk to patients if used.

    Product
    Piperacillin and Tazobactam for Injection, USP 4.5 grams per vial, Rx only, Mfd for Meitheal Pharmaceuticals Chicago, IL 60631, NDC 71288-004-51
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0579-2023·2023-05-24

    Heparin Sodium and Dextrose Injection Recalled Due to Subpotency

    B. Braun Medical Inc. is recalling 1,380 units of Heparin Sodium in Dextrose injection because the product is subpotent with lower anti-factor IIa potency than specified.

    Product
    HEPARIN SODIUM IN DEXTROSE — HEPARIN SODIUM IN DEXTROSE (HEPARIN SODIUM AND DEXTROSE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0601-2023·2023-05-24

    Ampicillin and Sulbactam Injectable Recalled for Sterility Assurance Defect

    The FDA has recalled Ampicillin and Sulbactam for Injection due to lack of assured sterility. Approximately 119,340 vials distributed nationwide within the United States are affected.

    Product
    Ampicillin and Sulbactam for Injection, USP 1.5 grams per vial, Rx only, Manufactured for: Xellia Pharmaceuticals USA, LLC, Buffalo Grove, IL 60089, NDC 70594-081-02
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0580-2023·2023-05-24

    Ampicillin for Injection Recall Due to Sterility Assurance Concerns

    Astral SteriTech is recalling 70,040 vials of Ampicillin for Injection nationwide due to lack of assurance of sterility in the manufacturing process.

    Product
    Ampicillin for Injection, USP 250 mg per vial, Rx only, Mfg. for: Piramal Critical Care, Bethlehem, PA 18017 USA, NDC 66794-220-41
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0627-2023·2023-05-24

    Piperacillin and Tazobactam Injection Recalled for Lack of Sterility Assurance

    Astral SteriTech recalls 223,590 vials of Piperacillin and Tazobactam for Injection nationwide due to lack of assured sterility. Patients and healthcare providers should stop use and consult their provider.

    Product
    Piperacillin and Tazobactam for Injection, USP, 3.375 grams per vial, Rx only, Mfg. for: Piramal Critical Care, Bethlehem, PA 18017, NDC 66794-217-41
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1585-2023·2023-05-24

    Philips Incisive CT system foot switch poses entrapment risk during unload

    The Philips Incisive CT system's foot switch may trap the operator's foot during the unload function, causing injury. The FDA has issued a Class II recall affecting 46 units worldwide.

    Product
    Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72814
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0619-2023·2023-05-24

    Injectable Ceftriaxone Recalled for Lack of Sterility Assurance

    Astral SteriTech is recalling 243,825 vials of Ceftriaxone for Injection due to inability to assure sterility. The medication was distributed nationwide and poses a contamination risk.

    Product
    Ceftriaxone for Injection, USP, 2 grams per vial, Rx only, Manufactured for: Xellia Pharmaceuticals USA, LLC, Buffalo Grove, IL 60089, NDC 70594-097-02
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1599-2023·2023-05-24

    Urine Analysis Diagnostic Device Software Issue Prevents Carryover Detection

    Beckman Coulter is recalling UA WIN 10 diagnostic kits because an optional software flag for detecting sample carryover was enabled but not displayed, preventing proper investigation of potential test contamination.

    Product
    UA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Food)·F-0894-2023·2023-05-24

    Trader Joe's Genova Pesto Recalled for Undisclosed Milk and Walnut Allergens

    Trader Joe's Genova Pesto is recalled due to labeling error: the product contains milk and walnuts but the label does not disclose these allergens. Affected product was distributed in Arizona, California, Florida, and Washington.

    Product
    Top Label: TRADER JOE'S GENOVA PESTO NET WT. 7 OZ. (198g) Side Label: TRADER JOE'S HUMMUS DIP NET WT. 7 OZ (98g) CONTAINS: SESAME. DIST. & SOLD EXCLUSIVELY BY: TRADER JOE'S, MONROVIA, CA 91016 UPC 0015 7353
    Category
    Food
    Distribution
    4 states
  • HighFDA (Drugs)·D-0642-2023·2023-05-24

    Vancomycin Injectable Solution Recalled Due to Lack of Sterility Assurance

    Apollo Care, LLC is recalling 413 bags of Vancomycin 2g injectable solution (Lot #AC-016638, expiration 05/27/2023) distributed in Missouri due to lack of assurance of sterility.

    Product
    VANComycin 2g added to 500 mL of 0.9% Sodium Chloride Injection, Rx Only. Apollo Care, LLC, 3801 Mojave Ct., Suite 101, Columbia, MO 65202. NDC 71170-204-50
    Category
    Drug
    Distribution
    1 state
  • HighFDA (Drugs)·D-0616-2023·2023-05-24

    Ceftriaxone Injectable Antibiotic Recalled Nationwide for Sterility Assurance Deficiency

    Astral SteriTech's 66,125-vial ceftriaxone injection is being recalled nationwide due to lack of assurance of sterility in manufacturing. The product is distributed across the United States.

    Product
    Ceftriaxone for Injection, USP 250 mg per vial, Rx only, Manufactured for: Xellia Pharmaceuticals USA, LLC, Buffalo Grove, IL 60089, NDC 70594-094-02
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0624-2023·2023-05-24

    Piperacillin and Tazobactam for Injection Recalled Due to Sterility Assurance Failure

    Astral SteriTech recalls approximately 893,950 vials of piperacillin and tazobactam injection nationwide because the manufacturer cannot assure the product was properly sterilized.

    Product
    Piperacillin and Tazobactam for Injection, USP 3.375 grams per vial, Rx only, Mfd for Meitheal Pharmaceuticals Chicago, IL 60631, NDC 71288-003-31
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0611-2023·2023-05-24

    Ceftriaxone injection vials recalled for lack of sterility assurance

    Astral SteriTech is recalling approximately 66,250 vials of Ceftriaxone for Injection due to lack of sterility assurance. The injectable antibiotic poses a potential contamination risk if used.

    Product
    Ceftriaxone for Injection, USP 250 mg per vial, Rx only, Mfg. for: Piramal Critical Care, Bethlehem, PA 18017, NDC 66794-211-42
    Category
    Drug
    Distribution
    Distributed nationwide