Anesthesia Ventilator Breathing Circuit Connections May Detach During Use
Draeger Medical is recalling 130 units of the VentStar Anesthesia WT ventilator due to a defect where glued breathing circuit connections may loosen or detach during ventilation, potentially compromising anesthesia delivery.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: This FDA Class I recall involves a structural defect in a critical anesthesia delivery device. Although no illnesses or injuries have been reported, the potential for serious patient harm from improper anesthesia delivery during surgery warrants the Class I designation and minimum score of 4.
Plain-English summary
Draeger Medical, Inc. is recalling 130 units of the VentStar Anesthesia WT (P)180, MP00374 anesthesia ventilator distributed nationwide in the United States.
The ventilators contain a defect in which glued connections of the breathing circuit can become loose before or during ventilation, resulting in partial or complete detachment of circuit components.
A loose or detached breathing circuit connection could impair or prevent the delivery of anesthesia during surgical procedures. This creates a risk of serious patient harm.
Healthcare facilities using this device should discontinue use and contact Draeger Medical, Inc. for repair, replacement, or return instructions.
The recalled product
- Product
- VentStar Anesthesia WT (P)180, MP00374
- Manufacturer
- Draeger Medical, Inc.
- Affected units
- 130
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI-DI: 04048675421948
- All lots.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 22 recalled by Draeger Medical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 22 products in this recall →Related recalls
Same manufacturer · Draeger Medical, Inc.
See all →- SevereFDA Recalls Draeger Atlan A300XL Anesthesia Workstation for Ventilator Failure
FDA (Devices) · 2024-12-25
- SevereDraeger Atlan A300 anesthesia workstation ventilator failure FDA Class I recall
FDA (Devices) · 2024-12-25
- SevereAnesthesia workstation piston ventilator may fail before or during use
FDA (Devices) · 2024-11-20
- CriticalDraeger Atlan A350 Anesthesia Workstation Piston Ventilator Failure
FDA (Devices) · 2024-11-20
- SevereDräger Perseus A500 Anesthesia Workstation Recalled for Unexpected Battery-Power Shutdowns
FDA (Devices) · 2024-04-24
Same hazard · connection failure
See all →- HighRapid Refill Continuous Injection System Recalled for Syringe Loosening Risk
FDA (Devices) · 2026-07-22
- HighBD Connecta stopcock Port C may over thread rather than tightening
FDA (Devices) · 2026-06-24
- SevereHalyard CATH LAB Kit Due to Syringe Adapter Disconnection Risk
FDA (Devices) · 2026-05-13
- SevereHalyard SAMMC Angiography Kit syringe rotating adapter unwinding recall
FDA (Devices) · 2026-05-13
- SevereHalyard Cardiac Catheterization Tray Kits Recalled for Connection Risk
FDA (Devices) · 2026-05-13