Steris Reliance Vision Washer/Disinfector Recalled for Fire and Smoke Risk
Steris is recalling 322 Reliance Vision Multi-Chamber Washer/Disinfectors because an electrical contactor malfunction could cause heating elements to overheat and produce smoke or fire.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II medical device recall involving a potential fire and smoke hazard. No injuries, fires, or property damage have been reported; the hazard is theoretical and described as a 'remote occurrence.' Per the severity rubric, risk-of-harm products with injury not yet reported score 3 (High).
Plain-English summary
Steris Corporation is recalling the Reliance Vision Multi-Chamber Washer/Disinfector (Reference Number FH83062) due to a potential electrical hazard in the drying chamber.
If the electrical contactor component in the drying chamber malfunctions, the heating elements could overheat, potentially resulting in smoke and/or fire.
Approximately 322 units have been distributed worldwide, including to customers in the United States, China, Lithuania, Saudi Arabia, South Korea, Canada, Taiwan, the United Kingdom, Spain, Australia, New Zealand, Brazil, Bahrain, Switzerland, South Africa, Hong Kong, and Qatar. Consumers and healthcare facilities with affected units should contact Steris Corporation for guidance.
The recalled product
- Product
- Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062
- Manufacturer
- Steris Corporation
- Hazard
- fire
- smoke
- electrical-malfunction
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI: 00724995193164
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighPhilips Azurion X-Ray Systems Table Movement Control Defect Recall
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula cardiopulmonary bypass catheter
FDA (Devices) · 2026-05-27
- HighTangent Single Use Digital Catheter Recalled for Distal Shaft Fractures
FDA (Devices) · 2026-05-27