Abbott Alinity m HPV test kit control failures from iron contamination
Abbott is recalling 13,318 units of its Alinity m HPV AMP Kit due to iron leaching into the lysis solution, causing positive controls to fail and potentially producing false negative test results.
- Product
- Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
- Category
- Medical Device
- Distribution
- Distributed nationwide