GE Optima MR750 MRI Scanner Recalled for Potential Shock Hazard
GE Medical Systems is recalling 68 Optima MR750 MRI scanners due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (shock hazard) where injury has not yet been reported, fitting the 'High' severity criteria for theoretical hazards without confirmed illness or injury.
Plain-English summary
GE Medical Systems, LLC is recalling 68 units of the Optima MR750 3.0T magnetic resonance (MR) scanner. The recall is due to a possible incorrect wiring of the Uninterruptable Power Supply (UPS), which could contribute to a potential shock hazard.
The affected units were distributed nationwide in the states of AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI. The product is a whole body magnetic resonance scanner designed for diagnostic imaging.
Healthcare facilities should contact GE Medical Systems to arrange for inspection and correction of the wiring issue to eliminate the shock hazard.
The recalled product
- Product
- Optima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal…
- Manufacturer
- GE Medical Systems, LLC
- Affected units
- 68
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Medical Device Listing No. D117354
- D295466
- D182389
- D102432
- D206140
- D230310
- D269569
- D102430
- D140572
- D188736
- D102434
- D234841
- D196247
- D134019
- D236543
- D006275
- D159181
- D288890
- D229859
- Model No. Eaton 9130
Distribution
Part of a larger recall action
This product is one of 9 recalled by GE Medical Systems, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · GE Medical Systems, LLC
See all →- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- HighGE Medical Optima XR200/220/240amx Mobile Radiographic Systems Recall
FDA (Devices) · 2026-07-22
- HighGiraffe OmniBed incubator canopy fasteners can fall into patient compartment
FDA (Devices) · 2026-06-24
- HighGiraffe OmniBed replaceable units share the canopy fastener disengagement defect
FDA (Devices) · 2026-06-24
Same hazard · shock hazard
See all →- High
- High
- High
- High
- Moderate
Similar recalls
- HighGE Healthcare Recalls MRI Scanners Due to Potential SAR Limit Exceedance
FDA (Devices) · 2015-11-18
- HighGE Healthcare Recalls 3.0T MR750 MRI Scanners Due to Coil Overheating Risk
FDA (Devices) · 2018-10-31
- ModerateGE Healthcare Recalls MR Scanners Due to Exposed Foam on End Bells
FDA (Devices) · 2013-10-09
- SevereGE Healthcare Recalls MR Scanners Due to Gradient Coil Overheating Risk
FDA (Devices) · 2014-04-30
- SevereGE Healthcare Recalls Discovery MR450 MRI Scanners Due to Burn Risk
FDA (Devices) · 2017-02-22
- HighGE Healthcare Recalls MRI Scanners Due to Cradle Release Malfunction
FDA (Devices) · 2014-05-07