FDA (Devices) recall action 77488

GE Medical Systems, LLC recalled 9 products on 2017-08-23

The FDA (Devices) publishes a recall like this one as 9 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 9 pages.

Products
9
Announced
2017-08-23
Critical
0
Units
612

Why these products were recalled

Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–9 of 9

  • HighFDA (Devices)··2017-08-23

    GE MRI Scanners Recalled for Potential Power Supply Shock Hazard

    GE Medical Systems is recalling 68 MRI scanners due to possible incorrect wiring in the uninterruptible power supply, which could create a shock hazard.

    Product
    1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Medical Systems Recalls Optima MR450 1.5T Scanners for Wiring Defect

    GE Medical Systems is recalling 68 Optima MR450 1.5T MRI scanners due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.

    Product
    Optima MR450 1.5T Product Usage: 1 .5T Brivo MR355 and 1.5T Optima MR360 are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Medical Systems Recalls Discovery MR750 3.0T Scanners for Shock Hazard

    GE Medical Systems is recalling 68 Discovery MR750 3.0T MRI scanners due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.

    Product
    Discovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Medical Systems Recalls MRI Scanners Due to Power Supply Wiring

    GE Medical Systems is recalling 68 MRI scanners due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.

    Product
    SIGNA Creator, SIGNA ExplorerProduct Usage: 1.5T SIGNA Creator and 1.5T SIGNA Explorer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Brivo and Optima MRI Scanners Recalled for Power Supply Wiring

    GE Medical Systems is recalling 68 Brivo MR355 and Optima MR360 MRI scanners due to possible incorrect wiring in the uninterruptible power supply, which could create a shock hazard.

    Product
    Brivo MR355, Optima MR360 Product Usage: The Brivo MR355/Optima MR360 is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Optima MR450w 1.5T MRI Scanner Recalled for Potential Shock Hazard

    GE Medical Systems is recalling 68 Optima MR450w 1.5T MRI scanners due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.

    Product
    Optima MR450w 1.5T Product Usage: The Optima MR450w 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Signa MRI Systems Recalled for Potential Shock Hazard

    GE Medical Systems is recalling 68 Signa MRI systems due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.

    Product
    1.5T Signa HDx, 3.0T Signa HDx, 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The GE Signa¿ HDx MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa¿…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Medical Systems Recalls Signa PET/MR Systems Due to Shock Hazard

    GE Medical Systems is recalling 68 Signa PET/MR systems because possible incorrect wiring of the Uninterruptable Power Supply could contribute to a potential shock hazard.

    Product
    Signa PET/MR (K163619, K142098) Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired…
    Category
    Distribution
    23 states
  • HighFDA (Devices)··2017-08-23

    GE Optima MR750 MRI Scanner Recalled for Potential Shock Hazard

    GE Medical Systems is recalling 68 Optima MR750 MRI scanners due to possible incorrect wiring in the Uninterruptable Power Supply, which could create a shock hazard.

    Product
    Optima MR750 3.0T Product Usage: The Optima MR450 1.5T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal…
    Category
    Distribution
    23 states