GE MRI Scanners Recalled for Potential Power Supply Shock Hazard
GE Medical Systems is recalling 68 MRI scanners due to possible incorrect wiring in the uninterruptible power supply, which could create a shock hazard.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (shock hazard) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
GE Medical Systems, LLC is recalling 68 units of 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, and Signa Vibrant magnetic resonance imaging (MRI) scanners. The recall is due to a possible incorrect wiring of the Uninterruptable Power Supply (UPS), which could contribute to a potential shock hazard.
The affected units were distributed nationwide in the United States, specifically in the states of AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI. The FDA has classified this recall as Class II, indicating that use of, or exposure to, the violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Healthcare facilities and users of these MRI scanners should contact GE Medical Systems for instructions on how to resolve the wiring issue. The recall number is Z-3028-2017.
The recalled product
- Product
- 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated…
- Manufacturer
- GE Medical Systems, LLC
- Category
- Medical Device — MRI Scanner
- Affected units
- 68
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Medical Device Listing No. D117354
- D295466
- D182389
- D102432
- D206140
- D230310
- D269569
- D102430
- D140572
- D188736
- D102434
- D234841
- D196247
- D134019
- D236543
- D006275
- D159181
- D288890
- D229859
- Model No. Eaton 9130
Distribution
Part of a larger recall action
This product is one of 9 recalled by GE Medical Systems, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · GE Medical Systems, LLC
See all →- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- HighGE Medical Optima XR200/220/240amx Mobile Radiographic Systems Recall
FDA (Devices) · 2026-07-22
- HighGiraffe OmniBed incubator canopy fasteners can fall into patient compartment
FDA (Devices) · 2026-06-24
- HighGiraffe OmniBed replaceable units share the canopy fastener disengagement defect
FDA (Devices) · 2026-06-24
Same hazard · shock hazard
See all →- High
- High
- High
- Moderate
- HighLe Hao Tool pressure washers lack a ground-fault circuit interrupter
CPSC · 2026-04-16
Similar recalls
- HighGE Healthcare Recalls Signa HDxt 3.0T MRI Scanners for SAR Limit Concerns
FDA (Devices) · 2015-11-18
- ModerateGE Healthcare Recalls MRI Systems Due to Potential Date Setting Error
FDA (Devices) · 2020-02-05
- HighGE Healthcare Recalls Signa HDx 3.0T MRI Scanners for SAR Display Error
FDA (Devices) · 2015-11-18
- ModerateGE Healthcare Recalls MRI Systems Due to Reporting Tool Error
FDA (Devices) · 2013-03-20
- ModerateGE Healthcare Recalls MRI Scanners Due to Potential Image Orientation Errors
FDA (Devices) · 2015-08-12
- ModerateGE Healthcare Recalls MRI Scanners Due to Non-Unique Series Identifiers
FDA (Devices) · 2013-10-02