Senographe Pristina
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can…
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
Severity not yet assessed
We have not scored this recall against our rubric yet, so we are not showing a severity rating or our usual guidance. The agency’s own notice below is the authoritative source — read it first.
Plain-English summary
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
The recalled product
- Product
- Senographe Pristina
- Manufacturer
- GE Medical Systems, LLC
- Category
- Medical Device — Devices
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (11)
- GTIN(s): 00840682118460
- 00840682142052
- 00840682145879
- 00195278022745
- 00195278276971
- 00195278661999
- 00195278665379
- 00195278914026
- 00198953109281
- 10198953054311
- 10195278901764.
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · GE Medical Systems, LLC
See all →- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- HighGE Medical Optima XR200/220/240amx Mobile Radiographic Systems Recall
FDA (Devices) · 2026-07-22
- HighGiraffe OmniBed incubator canopy fasteners can fall into patient compartment
FDA (Devices) · 2026-06-24
- HighGiraffe OmniBed replaceable units share the canopy fastener disengagement defect
FDA (Devices) · 2026-06-24