GE Healthcare Recalls MR Scanners Due to Gradient Coil Overheating Risk
GE Healthcare is recalling 1,546 MRI scanners worldwide because gradient coil connections may loosen, causing overheating, smoke, and potential fire.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a risk of fire and smoke generation in a clinical setting, which constitutes a significant injury/property-damage risk under the rubric for Severe (Score 4).
Plain-English summary
GE Healthcare, LLC is recalling 1,546 units of Discovery MR450, Discovery MR750, Optima MR450w, Optima MR 450w GEM, Discovery MR750w, and Discovery MR750w GEM magnetic resonance diagnostic devices. The recall covers 704 units in the United States and 842 units outside the US, with distribution across numerous countries worldwide.
The recall is due to a potential safety issue with the gradient coil electromechanical connections. These connections may loosen, causing an increase in electrical resistance. The increased resistance can lead to increasing heat during scanning, which may cause material in the Rear Endbell enclosure to overheat. This overheating can generate smoke in the scan room and cause localized fire.
The affected devices are distributed nationwide in the US (excluding North Dakota and Rhode Island, including DC, GU, and PR) and in many international locations. No specific consumer action is detailed in the provided source text, but the hazard involves fire and smoke generation during operation.
The recalled product
- Product
- GE Healthcare, Discovery MR450, Discovery MR750, Optima MR450w, Optima MR 450w GEM, Discovery MR750w, and Discovery MR750w GEM products. Magnetic Resonance Diagnostic Device.
- Manufacturer
- GE Healthcare, LLC
- Hazard
Distribution
Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- SevereGE Healthcare Anesthesia Systems Recalled for Potential Transducer Connection Reversal
FDA (Devices) · 2022-07-27
Same hazard · fire
See all →- SevereVQQQ Gas Patio Heaters Recalled for Fire and Burn Hazards
CPSC · 2026-10-22
- SevereAudi Recalls 2024-2025 Q4 E-Tron Models Due to Battery Fire Risk
NHTSA · 2026-10-21
- HighAutocar Recalls Xpeditor Trucks for Fuel Hose Separation Risk
NHTSA · 2026-10-21
- HighForest River Recalls 2023-2026 Pause Travel Trailers for Wiring Defect
NHTSA · 2026-10-20
- SevereForest River Recalls 2026-2027 Cedar Creek Trailers for Fire Risk
NHTSA · 2026-10-16
Similar recalls
- HighGE Healthcare Recalls 3.0T MR750 MRI Scanners Due to Coil Overheating Risk
FDA (Devices) · 2018-10-31
- HighGE Healthcare Recalls MRI Scanners Due to Potential SAR Limit Exceedance
FDA (Devices) · 2015-11-18
- HighGE Healthcare Recalls Discovery MR750w MRI Scanners Due to Burn Risk
FDA (Devices) · 2016-07-13
- SevereGE Healthcare Recalls Discovery MR450 MRI Scanners Due to Burn Risk
FDA (Devices) · 2017-02-22
- ModerateGE Healthcare Recalls MR Scanners Due to Exposed Foam on End Bells
FDA (Devices) · 2013-10-09
- HighGE Healthcare Recalls MRI Scanners Due to Cradle Release Malfunction
FDA (Devices) · 2014-05-07