GE Healthcare CT, PET, and NM Patient Tables May Drop During Service
GE Healthcare patient tables used in CT, PET, and NM systems can drop during service if technicians remove incorrect screws without appropriate supports, potentially injuring service personnel.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall involving risk of serious injury (table drop hazard) to service personnel during maintenance activities. The source text does not report any documented injuries or hospitalizations, placing this at the 'risk-of-harm products where injury has not yet been reported' level per the rubric.
Plain-English summary
This recall affects service manuals for GE Healthcare patient tables used in CT, PET, and NM imaging systems, including models such as GT1700, GT2000, Discovery, Optima, Revolution, and others. Approximately 24,843 tables are subject to this recall. The devices were distributed nationwide in the United States and Puerto Rico, with additional distribution to U.S. government and military facilities and over 100 countries worldwide.
The hazard arises from unclear guidance in the service manuals. Due to this unclear guidance, service personnel performing maintenance may remove incorrect screws. If proper table supports are not in place during this maintenance activity, the table can drop unexpectedly, creating a risk of serious injury to service technicians.
GE Healthcare classified this as a Class II recall under FDA oversight. Healthcare facilities operating these systems should ensure that service personnel are aware of the hazard and follow proper maintenance procedures, including appropriate use of table supports, to prevent table drops.
The recalled product
- Product
- Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for…
- Manufacturer
- GE Healthcare, LLC
- Hazard
- Affected units
- 24,843
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Rev 23 and prior
- GTIN: N/A
- (2) Service manual identifier 5307450-2EN
- Rev 30 and prior
- GTIN: Revolution Discovery CT: 00840682102490
- Revolution Discovery CT A: 00840682125574
- Revolution HD: N/A
- Discovery CT750 HD: N/A
- Rev 28 and prior
- GTIN: Optima CT540: 00840682102551
- Rev 2 and prior
- GTIN: 00195278563408. (5) Service manual identifier 5366080-8EN
- Rev 39 and prior
- Rev 14 and prior
- GTIN: N/A. (7) Service manual identifier 5809950-8EN
- Rev 3 and prior
- GTIN: Revolution Maxima: 00840682146180
- Revolution Maxima US: 00840682147392
- Revolution Ace: 00840682146159
- Optima CT680 Expert: 008406821469372
Distribution
Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- HighGE Anesthesia Systems Base Cracking and Fall Hazard Recalled
FDA (Devices) · 2022-07-27
- SevereGE Healthcare Anesthesia Systems Recalled for Potential Transducer Connection Reversal
FDA (Devices) · 2022-07-27
Same hazard · drop hazard
See all →- HighLuminos dRF Max X-ray Display Support Arm May Unintentionally Lower
FDA (Devices) · 2024-08-14
- HighAeos Robotic Digital Microscope recalled for safety mechanism failure
FDA (Devices) · 2024-06-19
- HighCardiac SPECT System Recalled for Defective Service Handle Fastening
FDA (Devices) · 2024-06-19
- HighHeartland travel trailers recalled for retractable awning defect
NHTSA · 2022-08-08
- HighAlliance RV Paradigm and Valor Travel Trailers Awning Weld Separation
NHTSA · 2022-07-20
Similar recalls
- HighGE Healthcare Recalls Discovery NM/CT 670 and NM 630 Systems
FDA (Devices) · 2012-10-31
- HighGE Healthcare Recalls Discovery PET/CT 710 Systems Due to Table Stop Defect
FDA (Devices) · 2019-06-26
- SevereGE Healthcare Recalls Discovery ML DR PET/CT Systems Due to Table Stop Defect
FDA (Devices) · 2019-06-26
- HighGE Healthcare Nuclear Medicine Systems Detector Fall Risk
FDA (Devices) · 2025-01-22
- HighGE Revolution CT Patient Table Recalled for Unintended Motion Risk
FDA (Devices) · 2016-06-15
- HighGE Healthcare Recalls Discovery PET/CT 610 Systems Due to X-Ray Safety Issue
FDA (Devices) · 2019-06-26