GE Healthcare Recalls Discovery PET/CT 710 Systems Due to Table Stop Defect
GE Healthcare is recalling 1,113 Discovery PET/CT 710 systems because the table may stop moving without stopping X-ray exposure.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (excessive radiation exposure) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
GE Healthcare, LLC is recalling 1,113 Discovery PET/CT 710 systems distributed nationwide. The recall is due to a defect where the patient table may stop moving without stopping the X-ray exposure.
The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
The source text does not specify the corrective action or instructions for consumers.
The recalled product
- Product
- Discovery PET/CT 710 Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
- Manufacturer
- GE Healthcare, LLC
- Hazard
- Affected units
- 1,113
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- CT 710
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by GE Healthcare, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- HighSIGNA Premier MRI scanner recalled for elevated acoustic noise hazard
FDA (Devices) · 2022-07-27
Same hazard · radiation exposure
See all →- HighPhilips Ingenuity CT Scanners Recalled Due to Bolus Tracking Trigger Issue
FDA (Devices) · 2026-08-05
- HighPhilips CT Scanners Recalled Due to Bolus Tracking Trigger Issue
FDA (Devices) · 2026-08-05
- HighPhilips IQon Spectral CT Recalled Due to Automatic Scan Trigger Failure
FDA (Devices) · 2026-08-05
- HighPhilips Allura Xper hand switch buttons may not fully release
FDA (Devices) · 2026-06-24
- HighPhilips MultiDiagnost Eleva hand switch buttons may not fully release
FDA (Devices) · 2026-06-24