Agency classClass IIOur severity:HighFDA (Devices)·Z-2852-2018·Announced 2018-08-29

Philips DigitalDiagnost 4.1 Footplate Hook Recall for Injury Risk

Philips is recalling DigitalDiagnost 4.1 units because a defective hook allows the footplate to fall, posing a risk of injury to patients or operators.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product where injury has not yet been reported, fitting the criteria for a High severity score.

Plain-English summary

Philips Medical Systems Gmbh, DMC is recalling the DigitalDiagnost 4.1 (Stitching Patient Support) 712225. The recall is due to a defect in the hook that is designed to secure the footplate in a vertical position. Because the hook does not hold the footplate securely, the footplate can fall or swing down, which may result in injury to the operator or patient.

The affected product is identified by Serial Number SN16000137. The distribution pattern for this recall includes the US Nationwide and international distribution to 66 countries. The agency has classified this recall as Class II.

Consumers and healthcare facilities should stop using the affected DigitalDiagnost 4.1 units immediately. Users should contact the recalling firm, Philips Medical Systems Gmbh, DMC, for instructions on how to resolve the issue or obtain a replacement.

The recalled product

Product
DigitalDiagnost 4.1 (Stitching Patient Support) 712225

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Serial Number: SN16000137

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 28 recalled by Philips Medical Systems Gmbh, DMC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 28 products in this recall →

Same manufacturer · Philips Medical Systems Gmbh, DMC