Curbell Medical Recalls CareSense Bed and Chair Monitors

Curbell Medical is recalling CareSense Wireless Advanced LCD Bed and Chair Monitors because a circuit board modification disabled the alarm function.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device where a safety-critical alarm function has failed, creating a risk of harm to patients, although no specific injuries or illnesses are reported in the source text.

Plain-English summary

Curbell Medical, Inc. is recalling the CareSense Wireless Advanced LCD Bed and Chair Monitor, Model # CSM-BC400. This device is a bed patient monitor connected to a sensor pad that sounds an alarm if a patient attempts to arise from a bed, chair, floor mat, or toilet seat.

The recall was initiated after a customer complaint revealed that a resistor was incorrectly placed on the monitor's circuit board. This modification was a planned change intended to improve battery drain, but it resulted in the monitor's alarm via the nurse call communication port failing to function.

The recall affects 112 domestic units and 211 foreign units distributed nationwide, as well as in Canada, Ireland, and Italy. The affected units are identified by Lot 0414. Consumers and healthcare facilities should stop using the affected monitors and contact Curbell Medical, Inc. for instructions on how to proceed.

The recalled product

Product
CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a…

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot 0414

Distribution

Distributed nationwide across the United States.