Spacelabs Xprezzon Bedside Monitor Recalled for Display Disconnection
Spacelabs Healthcare is recalling 21 Xprezzon Bedside Monitors because specific key presses can cause patient parameters to disappear from the display.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where a software defect causes patient parameters to disappear from the display, creating a risk of harm due to lack of monitoring, although no injuries or illnesses have been reported.
Plain-English summary
Spacelabs Healthcare, LLC is recalling 21 units of the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. These patient monitors are intended to alert users to alarm conditions and display patient parameters in hospital environments. The recall affects units distributed in the United States (including Florida, Iowa, Oklahoma, and South Carolina) and in India, Germany, and Panama.
The recall was initiated due to reports that a specific combination of key presses can cause the module to disconnect from the monitor. When this occurs, patient parameters drop from the display. The parameters return to the display after the monitor is power cycled.
Healthcare facilities should identify affected units using the provided serial numbers and contact Spacelabs Healthcare for instructions on how to resolve the issue.
The recalled product
- Product
- Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs
- Manufacturer
- Spacelabs Healthcare, Llc
- Category
- Medical Device — Patient Monitor
- Affected units
- 21
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 040-1632-00
- 1393-000013
- 1393-000039
- 1393-000040
- 1393-000041
- 1393-000341
- 1393-000382
- 1393-000383
- 1393-000384
- 1393-000385
- 1393-000386
- 1393-000387
- 1393-000388
- 1393-000808
- 1393-000831
- 1393-001374
- 1393-001375
- 1393-001453
- 1393-001990
- 1393-001991
Distribution
Related recalls
Same manufacturer · Spacelabs Healthcare, Llc
See all →- ModerateSpacelabs Recalls G2 Clinical Access Software CDs Due to Incorrect Programs
FDA (Devices) · 2014-04-02
- HighSpacelabs AriaTelemetry Transmitters Recalled for SpO2 Display Errors
FDA (Devices) · 2014-02-05
- HighSpacelabs qube Patient Monitor Recalled for Power-On Failure
FDA (Devices) · 2013-11-27
- ModerateSpacelabs elance Vital Signs Monitor Recalled Due to Sensor Connection Failure
FDA (Devices) · 2013-11-20
- HighSpacelabs Smart Disclosure System Recalled for Incorrect Waveform Display
FDA (Devices) · 2013-01-23
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25