United Imaging uCT 550 CT Systems Recalled for Software Defect
Shanghai United Imaging Healthcare is recalling 23 uCT 550 CT systems due to a software issue that may hide patient images, potentially requiring rescans.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a software defect that may lead to rescans rather than direct injury or death, fitting the criteria for a moderate severity level.
Plain-English summary
Shanghai United Imaging Healthcare Co., Ltd. is recalling 23 units of the Computed Tomography X-ray System, Model uCT 550, due to a software defect. The recall covers systems distributed nationwide in the United States, including specific states such as California, Texas, and New York.
The software issue occurs when the Split or Local function is enabled and multiple RIS exams are merged into a single study. This defect prevents patient images from being visible on the Patient Administration interface. As a result, healthcare facilities may need to perform rescans of patients to obtain the necessary images.
Healthcare facilities should contact Shanghai United Imaging Healthcare Co., Ltd. for instructions on how to resolve the software issue. The recall is classified as Class II by the FDA, indicating that the defect may cause temporary or medically reversible health consequences.
The recalled product
- Product
- Computed Tomography X-ray System, Model: uCT 550 with software
- Manufacturer
- Shanghai United Imaging Healthcare Co., Ltd.
- Affected units
- 23
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 353045/R001.19.6.1857994-Re3300SP1H1_P5
- 353049/R001.19.6.1857994-Re3300SP1H1_P5
- 354001/R001.19.6.1857994-Re3300SP1H1_P5
- 354004/R001.19.6.1857994-Re3300SP1H1_P5
- 354005/R001.19.6.1857994-Re3300SP1H1_P5
- 354007/R001.19.6.1857994-Re3300SP1H1_P5
- 354009/R001.19.6.1857994-Re3300SP1H1_P5
- 354013/R001.19.6.1857994-Re3300SP1H1_P5
- 354015/R001.19.6.1857994-Re3300SP1H1_P5
- 354016/R001.19.6.1857994-Re3300SP1H1_P5
- 354018/R001.17.6.1607336-Re3100SP4
- 354020/R001.17.6.1607336-Re3100SP4
- 354021/R001.17.6.1607336-Re3100SP4
- 354022/R001.19.4.1810621-Re3300SP1H1_P3
- 354023/R001.19.6.1857994-Re3300SP1H1_P5
- 354024/R001.19.6.1857994-Re3300SP1H1_P5
- 354025/R001.19.6.1857994-Re3300SP1H1_P5
- 354026/R001.19.6.1857994-Re3300SP1H1_P5
- 354027/R001.19.6.1857994-Re3300SP1H1_P5
- 354028/R001.19.6.1857994-Re3300SP1H1_P5
Distribution
Related recalls
Same manufacturer · Shanghai United Imaging Healthcare Co., Ltd.
See all →- ModerateUnited Imaging uMI Panorama GS PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uMI Panorama 35 PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging CT System Recall Due to Software Image Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uCT 760 CT System Recalled for Software Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uMI 550 PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
Same hazard · software defect
See all →- SevereWaymo Recalls Automated Driving Systems for Closed Freeway Lane Risk
NHTSA · 2026-11-20
- High
- SevereRam Recalls 2022-2024 1500 Classic Trucks for ABS Software Defect
NHTSA · 2026-11-19
- ModeratePhilips Diagnostic Ultrasound Systems Recalled for Software Measurement Errors
FDA (Devices) · 2026-11-18
- ModerateSiemens ACUSON Sequoia Ultrasound System Recalled for Software Issue
FDA (Devices) · 2026-11-18
Similar recalls
- HighUnited Imaging uCT 530 CT Systems Recalled for Software Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uCT 780 CT System Recalled for Software Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uCT ATLAS CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging CT System Recall Due to Software Image Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uCT 760 CT System Recalled for Software Visibility Issue
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uMI 550 PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07