Agency classClass IIOur severity:HighFDA (Devices)·Z-3277-2026·Announced 2026-10-07

United Imaging uCT 530 CT Systems Recalled for Software Visibility Issue

Shanghai United Imaging Healthcare is recalling four uCT 530 CT systems due to a software defect that may hide patient images, potentially requiring rescans.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of harm (need for rescans) without reported injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Shanghai United Imaging Healthcare Co., Ltd. is recalling four units of the Computed Tomography X-ray System, Model uCT 530, due to a software defect. The recall covers systems distributed nationwide in specific US states, including California, Texas, and New York.

The software issue occurs when the Split or Local function is enabled and multiple RIS exams are merged into a single study. This defect prevents patient images from being visible on the Patient Administration interface. As a result, healthcare facilities may need to perform rescans of patients to ensure images are properly captured and available.

Healthcare providers should contact Shanghai United Imaging Healthcare to resolve the software issue. The recall is classified as Class II by the FDA, indicating that the defect may cause temporary or medically reversible health consequences, such as the need for additional imaging procedures.

The recalled product

Product
Computed Tomography X-ray System, Model: uCT 530 with software
Affected units
4

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • 304006/R001.17.6.1607336-Re3100SP4
  • 304007/R001.19.4.1810621-Re3300SP1H1_P3
  • 304008/R001.19.4.1810621-Re3300SP1H1_P3

Distribution

Distributed nationwide across the United States.