The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2114-2014·Announced 2014-07-30

GE Healthcare Recalls CARESCAPE Monitor B650 Due to Sound Loss

GE Healthcare is recalling CARESCAPE Monitor B650 units because they may lose alarm sounds when an external display is connected via a USB extender with an active hub.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device where the loss of alarm sounds poses a risk of harm to patients, but no injuries or illnesses have been reported, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

GE Healthcare, LLC is recalling the CARESCAPE Monitor B650, a multi-parameter patient monitor used in professional healthcare facilities. The recall involves 21,799 units distributed worldwide, including 7,635 units in the United States and 14,164 units outside the US.

The recall is due to a potential sound loss issue. When an external display is connected to the monitor using a USB extender cable with an active USB hub, the system may lose sound, including critical alarm sounds, if the external display is powered off and on during normal operation.

Healthcare facilities should check their inventory for affected serial numbers listed in the recall notice. If a unit is affected, the facility should contact GE Healthcare for instructions on corrective action to ensure alarm functionality is restored.

The recalled product

Product
GE Healthcare, CARESCAPE Monitor B650. A multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.
Manufacturer
GE Healthcare, LLC

Distribution

Distributed nationwide across the United States.

The notice also names 2 states specifically:

Similar recalls