The Recall Desk

Manufacturer

Curbell Medical, Inc.

1 recall in our database name Curbell Medical, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
1
Critical
0
Severe
0
Most recent
2014-09-03

Of the 1 recalls from Curbell Medical, Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–1 of 1

  • HighFDA (Devices)·Z-2554-2014·2014-09-03

    Curbell Medical Recalls CareSense Bed and Chair Monitors

    Curbell Medical is recalling CareSense Wireless Advanced LCD Bed and Chair Monitors because a circuit board modification disabled the alarm function.

    Product
    CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed
    Category
    Medical Device
    Distribution
    Distributed nationwide