RaySearch TomoTherapy SharePlan Software Field Notice for Unexpected Behaviors
RaySearch Laboratories issued a field notice for TomoTherapy SharePlan software due to two unexpected behaviors not clearly described in the user manual. No patient mistreatment or incidents have occurred.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a medical device software where specific user actions are required to ensure clinical outcomes, representing a risk of harm where injury has not yet been reported.
Plain-English summary
RaySearch Laboratories AB has issued a field notice for the TomoTherapy SharePlan 1.1 software, also known as t-RayAutoplan version 1.2.1 (build number 1.3.1.10). The notice addresses two specific behaviors of the software, involving Multiple Fraction Groups and Depth Offset, that may be unexpected to users and are not clearly described in the user manual.
The agency classifies this as a Class II recall. According to the source text, none of these behaviors have caused patient mistreatment or other incidents. However, the notice states that specific user actions must be adhered to for the best agreement between treatment plan evaluation and the clinical outcome of the delivered treatment.
The affected software was distributed worldwide, including in the United States (California, Kentucky, Michigan, New York, Oklahoma, and Oregon) and internationally to Canada, Belgium, the United Kingdom, Australia, France, Germany, and Poland. A total of 15 units are affected. Users should check the About box in the SharePlan application to verify if the build number is 1.3.1.10.
The recalled product
- Product
- The product affected by this notice is sold under the trade name TomoTherapy SharePlan 1.1. To determine if the version you are using is affected, bring up the About box in the SharePlan application and check if the build number reported there is 1.3.1.10. Intended use…
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Category
- Medical Device — Software
- Affected units
- 15
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- SharePlan 1.1
- aka t-RayAutoplan version 1.2.1 (build number 1.3.1.10).
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software error
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- HighPhilips Spectral CT 7500 Software Recall for Radiation Exposure Risk
FDA (Devices) · 2026-10-07
- HighBard EnCor EnCompass Breast Biopsy System Software Recall
FDA (Devices) · 2026-10-07
- ModerateBeckman Coulter AU480 Chemistry Analyzer Software Error Recall
FDA (Devices) · 2026-10-07
- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
Similar recalls
- ModerateRaySearch SharePlan Software Recalled for Unexpected Import Behavior
FDA (Devices) · 2012-10-03
- SevereRaySearch Ray Station Software Recalled for Dose Miscalculations
FDA (Devices) · 2014-04-30
- HighRayStation software recall for potential incorrect treatment plan references
FDA (Devices) · 2020-12-02
- HighRayStation Radiation Therapy Software Recalled for Potential Dose Errors
FDA (Devices) · 2014-04-09
- HighRayStation Treatment Planning Software Recalled for Dose Calculation Error
FDA (Devices) · 2012-09-12
- HighTomoTherapy Hi-Art Software Recall for Potential Safety Anomalies
FDA (Devices) · 2014-08-27