RayStation software recall for potential incorrect treatment plan references
RaySearch Laboratories is recalling RayStation software because missing data may cause treatment plans to reference incorrect regions or points of interest.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device software defect that could lead to incorrect treatment planning, representing a risk of harm where injury has not yet been reported, consistent with the 'High' severity criteria.
Plain-English summary
RaySearch Laboratories AB is recalling RayStation standalone software treatment planning systems, including models 4.0 through 10.0. The recall involves 2,768 units distributed nationwide in the United States.
The recall is due to a software defect where a region of interest (ROI) or point of interest (POI) referenced from an imported plan may become linked to the wrong ROI or POI if that specific data is missing in the imported RT Structure Set. This could result in incorrect treatment plan references.
Authorized intended users should contact RaySearch Laboratories for instructions on how to resolve the issue. The source text does not report any specific injuries or illnesses associated with this defect.
The recalled product
- Product
- RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation propose
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Affected units
- 2,768
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Software version numbers: 4.0.0.14
- 4.0.3.4
- 4.5.1.14
- 4.7.2.5
- 4.7.3.13
- 4.7.4.4
- 4.7.5.4
- 4.7.6.7
- 4.9.0.42
- 5.0.1.11
- 5.0.2.35
- 5.0.3.17
- 6.0.0.24
- 6.1.1.2
- 6.2.0.7
- 6.3.0.6
- 7.0.0.19
- 8.0.0.61
- 8.0.1.10
- 8.1.0.47
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25