RayStation Software Recalled for Potential Radiation Overdose Risk
RaySearch Laboratories is recalling RayStation treatment planning software versions 4.5 through 8A because a calculation error could lead to significant radiation overdosage during therapy.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall involves a medical device defect that could lead to significant radiation overdosage, which constitutes a risk of serious injury, aligning with the 'Severe' criteria for significant injury reports or high-risk medical device defects.
Plain-English summary
RaySearch Laboratories AB is recalling 746 units of RayStation stand-alone software treatment planning systems, including versions 4.5, 4.7, 4.9, 5, 6, 7, and 8A. The software is used by clinically qualified radiation therapy staff to plan and analyze radiation treatments. The affected units were distributed nationwide to accounts in 33 states and the District of Columbia.
The recall is due to a defect in how the software calculates DMLC fields when the collimator is rotated and the beam model has a highly asymmetric primary source. This calculation error could lead to potentially significant overdosage at delivery. The source text notes that users must be aware of the issue to avoid incorrect dose calculations during treatment planning.
No specific dates, lot numbers, or reported injuries are provided in the source text. Users of the affected software versions should contact RaySearch Laboratories for further instructions on how to address the calculation issue.
The recalled product
- Product
- RayStation stand-alone software treatment planning system, RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) and RayStation 8A (RayPlan 8A) Product Usage: RayStation is a software system designed for
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Affected units
- 746
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (19)
- Software Versions 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.3.13
- 4.7.4.4
- 4.7.5.4
- 4.9.0.42
- 5.0.0.37
- 5.0.1.11
- 5.0.2.3
- 6.0.0.24
- 6.1.0.26
- 6.1.1.2
- 6.2.0.7
- 7.0.0.19
- 8.0.0.61
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · radiation overdose
See all →- HighX-Ray System Recalled for Potential Excessive Radiation Dose in First Image
FDA (Devices) · 2025-02-05
- HighARTIS Angiography Systems Recalled for Potential Radiation Overdose
FDA (Devices) · 2024-07-10
- HighSiemens ARTIS Angiography Systems May Deliver Unintended Radiation Doses
FDA (Devices) · 2024-07-10
- HighRadiation therapy planning system may report overly high skin surface doses
FDA (Devices) · 2024-05-08
- HighProton Therapy System Delivers Incorrect Radiation Dose in First Layer
FDA (Devices) · 2022-06-08