The Recall Desk

Hazard

Radiation Overdose recalls

14 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
14
Critical
0
Severe
7
Most recent
2025-02-05

Of the 14 radiation overdose recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 7 (50%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all radiation overdose recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–14 of 14

  • HighFDA (Devices)·Z-0922-2025·2025-02-05

    X-Ray System Recalled for Potential Excessive Radiation Dose in First Image

    Siemens ARTIS One X-ray systems may deliver higher radiation doses during first imaging after patient registration due to copper prefiltration malfunction occurring in very rare situations.

    Product
    ARTIS One Angiographic X-Ray System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2242-2024·2024-07-10

    ARTIS Angiography Systems Recalled for Potential Radiation Overdose

    Siemens ARTIS icono angiography systems with specific detector configurations may deliver higher radiation doses than intended in rare cases. The issue occurs during the first X-ray exposure after patient registration.

    Product
    ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2241-2024·2024-07-10

    Siemens ARTIS Angiography Systems May Deliver Unintended Radiation Doses

    Siemens ARTIS icono angiography systems with certain detectors may deliver higher-than-intended radiation doses in rare cases. The issue occurs when incorrect copper prefiltration is applied during initial x-ray exposure after patient registration.

    Product
    ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1742-2024·2024-05-08

    Radiation therapy planning system may report overly high skin surface doses

    RayStation versions 8.0.0.61 and 8.0.1.10 are subject to a Class II FDA recall due to a potential for calculated skin surface doses to be reported as too high. Healthcare facilities using these versions should contact the manufacturer for guidance and a corrective update.

    Product
    RayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1205-2022·2022-06-08

    Proton Therapy System Delivers Incorrect Radiation Dose in First Layer

    The Proteus235 Proton Therapy System may deliver radiation to initial treatment fields with incorrect scanning magnet settings. The radiation field may be smaller than expected and may deliver higher dose than prescribed.

    Product
    Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0867-2022·2022-04-13

    X-ray fluoroscopy system may deliver excessive radiation dose

    Shimadzu FLUOROspeed X1 fluoroscopy systems may deliver radiation doses exceeding federal regulatory limits in specific fluoroscopic operating modes due to inadequate installation adjustment criteria.

    Product
    MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
    Category
    Medical Device
    Distribution
    37 states
  • HighFDA (Devices)·Z-0868-2022·2022-04-13

    X-ray System Models Recalled for Potential Radiation Dose Exceeding Standards

    Shimadzu Medical Systems recalls 176 SonialVision G4 X-ray systems that may emit radiation doses exceeding federal standards in some fluoroscopic mode cases. No injuries have been reported.

    Product
    MODEL: X-RAY TV SYSTEM SONIALVISION G4
    Category
    Medical Device
    Distribution
    37 states
  • SevereFDA (Devices)·Z-1678-2020·2020-04-15

    Siemens Medical Solutions Recalls MEVATRON Radiation Therapy Systems

    Siemens Medical Solutions is recalling specific MEVATRON radiation therapy systems due to a potential safety issue with dose monitoring interlocks that could result in a 10-20% radiation overdose.

    Product
    MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #01940753, MEVATRON M-2 6740 #09401506, MEVATRON MD-2 #09401654, MEVATRON MDX-2 #09401746 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
    Category
    Medical Device
    Distribution
    35 states
  • SevereFDA (Devices)·Z-1679-2020·2020-04-15

    Siemens ONCOR Radiation Therapy Systems Recalled for Potential Dose Overdose

    Siemens is recalling specific ONCOR radiation therapy systems due to a potential safety issue with dose monitoring interlocks that could result in a 10-20% overdose.

    Product
    ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus #5857912 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
    Category
    Medical Device
    Distribution
    35 states
  • SevereFDA (Devices)·Z-1677-2020·2020-04-15

    Siemens ARTISTE MV System Recalled for Radiation Dose Monitoring Safety Interlock Defect

    Siemens is recalling ARTISTE MV Systems due to a potential safety issue with dose monitoring interlocks that could result in a 10-20% radiation therapy overdose.

    Product
    ARTISTE MV System Model #8139789 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
    Category
    Medical Device
    Distribution
    35 states
  • SevereFDA (Devices)·Z-1680-2020·2020-04-15

    Siemens PRIMUS HI Linear Accelerators Recalled for Radiation Overdose Risk

    Siemens is recalling PRIMUS HI linear accelerators because a safety interlock defect could cause a 10-20% overdose of radiation therapy.

    Product
    PRIMUS HI #04504200 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
    Category
    Medical Device
    Distribution
    35 states
  • SevereFDA (Devices)·Z-0821-2019·2019-02-20

    Elekta MOSAIQ Oncology System Recalled for Potential Radiation Dose Errors

    Elekta is recalling MOSAIQ Oncology Information Systems because missing wedge IDs may cause patients to receive higher radiation doses than planned.

    Product
    MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy a
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2497-2018·2018-07-25

    RayStation Software Recalled for Potential Radiation Overdose Risk

    RaySearch Laboratories is recalling RayStation treatment planning software versions 4.5 through 8A because a calculation error could lead to significant radiation overdosage during therapy.

    Product
    RayStation stand-alone software treatment planning system, RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) and RayStation 8A (RayPlan 8A) Product Usage: RayStation is a software system designed for
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0980-2015·2015-01-21

    Siemens Linear Accelerators Recalled for Unauthorized Component Safety Risks

    Siemens Medical Solutions is recalling 320 linear accelerators because unauthorized third-party components may cause patient or operator injury and dose calculation errors.

    Product
    Siemens Linear Accelerator: ARTISTE, ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIUS HI, Mevatron M2/Primus Mid-Energy, MEVATRON PRIMART, MEVATRON KD-2, MEVATRON KDS-2, MEVATRON K2, MEVATRON 6323-2, MEVATRON M2 6740, MEVATRON M2 6700, MEVATRON
    Category
    Medical Device
    Distribution
    Distributed nationwide