Beckman Coulter AU480 Chemistry Analyzer Software Error Recall
Beckman Coulter is recalling AU480 Chemistry Analyzers because a software error prevents the instrument from starting, delaying patient results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a delay in obtaining patient results due to software failure, which is a significant operational issue but does not involve direct physical injury or death.
Plain-English summary
Beckman Coulter Mishima K.K. is recalling certain AU480 Chemistry Analyzers. These devices are used for in vitro diagnostic testing to measure analytes in samples. The recall involves 492 units distributed worldwide, including in the United States and numerous other countries.
The recall is due to a software issue on the pre-installed Windows 10 IoT Enterprise 2021 LTSC PCs. When launching the instrument software, the device may display the message "System Stopped By Fatal Error: DBConnectionError AlarmDB." This error makes the AlarmDB unreadable and prevents the instrument software from starting. As a result, there is a delay in obtaining patient results.
Affected units are identified by specific item numbers and UDI-DI codes, including B96692, B96693, and B12183. Users should contact Beckman Coulter Mishima K.K. for instructions on how to resolve the software issue.
The recalled product
- Product
- Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate…
- Manufacturer
- Beckman Coulter Mishima K.K.
- Affected units
- 492
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- the UDI number is not available.)
Distribution
Related recalls
Same manufacturer · Beckman Coulter Mishima K.K.
See all →- HighBeckman Coulter AU5800 analyzer may stop analyzing after database errors
FDA (Devices) · 2026-03-18
- HighBeckman Coulter DxC 700 AU analyzer may stop analyzing after errors
FDA (Devices) · 2026-03-18
- HighBeckman Coulter Sample Probes with Defective Tips Clinical Chemistry
FDA (Devices) · 2025-06-11
- ModerateBeckman Coulter Sample Probe Manufacturing Defect Recall
FDA (Devices) · 2025-06-11
- ModerateClinical Chemistry Analyzer Software Issue Causes Delayed Results
FDA (Devices) · 2024-08-28
Same hazard · software error
See all →- ModerateCanon Alphenix Interventional Angiography System Recalled for Software Error
FDA (Devices) · 2026-11-04
- ModerateMedtronic CareLink SmartSync Device Manager App Software Recall
FDA (Devices) · 2026-11-04
- ModerateCanon Alphenix Interventional Angiography System Recalled for Software Error
FDA (Devices) · 2026-11-04
- ModerateBioMerieux EMAG System Recalled for Software Error Causing Delayed Results
FDA (Devices) · 2026-11-04
- ModerateCanon Alphenix Interventional Angiography System Software Recall
FDA (Devices) · 2026-11-04
Similar recalls
- HighBeckman Coulter AU5800 Clinical Chemistry Analyzer Recalled for Software Error
FDA (Devices) · 2013-05-01
- ModerateBeckman Coulter AU5800 Clinical Chemistry Analyzer Software Recall
FDA (Devices) · 2013-08-07
- HighBeckman Coulter AU5800 analyzer may stop analyzing after database errors
FDA (Devices) · 2026-03-18
- HighClinical Chemistry Analyzer Recalled for Software Defect in Test Result Calculation
FDA (Devices) · 2025-04-09
- HighBeckman Coulter DxC 500 AU Clinical Chemistry Analyzer Recall
FDA (Devices) · 2025-05-07
- ModerateBeckman Coulter DxC 500 AU Chemistry Analyzer Software Calibration Error
FDA (Devices) · 2025-05-21