Zimmer Biomet Recalls Vital MIS Spinal Fixation Implants Due to Locking Issues
Zimmer Biomet is recalling Vital MIS Spinal Fixation system implants because reports indicate the extended tab screws may not lock adequately, causing the closure top to back out post-operatively.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a structural failure of a medical implant (screw backing out) which poses a risk of harm, but the source text does not report specific injuries or hospitalizations.
Plain-English summary
Zimmer Biomet, Inc. is recalling 3,235 units of Vital MIS Spinal Fixation system implants, specifically the CANN POLY EXT TAB 4.5X40MM (Item Number 810M4540). These non-cervical spinal fixation devices are intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation in skeletally mature and adolescent patients.
The recall was initiated due to reports of inadequate locking of the extended tab screws. Complaints were received indicating that the closure top backed out of the screw post-operatively. The affected units were distributed nationwide in the United States.
Healthcare providers and patients should contact Zimmer Biomet for instructions on how to proceed with the recall. The affected lot numbers are C20B0006, SBM108401, SBM118983, SBM121852, and SBM122267. The UDI for the product is (01)00887868324208.
The recalled product
- Product
- Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 4.5X40MM, Item Number 810M4540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is…
- Manufacturer
- Zimmer Biomet, Inc.
- Affected units
- 3,235
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI: (01)00887868324208
- Lot Numbers: C20B0006 SBM108401 SBM118983 SBM121852 SBM122267
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 25 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 25 products in this recall →Related recalls
Same manufacturer · Zimmer Biomet, Inc.
See all →- HighZimmer Biomet Recalls Spinal Rod Cutters Due to Fracture Risk
FDA (Devices) · 2020-12-30
- ModerateZimmer Biomet Recalls Polaris Spinal System Screws for Labeling Update
FDA (Devices) · 2020-12-23
- HighZimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Concerns
FDA (Devices) · 2020-11-11
Same hazard · mechanical failure
See all →- SevereCanon Medical Alphenix INFX-8000C/B5 Recalled for Tabletop Shift Risk
FDA (Devices) · 2026-10-14
- ModerateLeMaitre TufTex Over-the-Wire Embolectomy Catheter Recalled for Balloon Failure
FDA (Devices) · 2026-10-14
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
Similar recalls
- ModerateZimmer Biomet Recalls Spinal Plates Due to Packaging Adhesion
FDA (Devices) · 2019-11-13
- HighZimmer Biomet Recalls Vitality Spinal Fixation Set Screws
FDA (Devices) · 2019-03-06
- HighBiomet Spine Polaris Screw Inserter Recalled for Tip Failure Risk
FDA (Devices) · 2013-03-27
- ModerateZimmer Biomet Recalls Ziptight Ankle Fixation Devices Due to Missing Tip Protectors
FDA (Devices) · 2018-03-07
- ModerateZimmer Biomet Recalls Distal Lateral Femoral Locking Plates
FDA (Devices) · 2019-11-13
- HighZimmer Biomet Recalls Ball Tip Guide Wires Due to Sterile Barrier Failure
FDA (Devices) · 2018-06-27