Agency classClass IIOur severity:HighFDA (Devices)·Z-2212-2020·Announced 2020-06-10

Zimmer Biomet Recalls Vital MIS Spinal Fixation Implants Due to Locking Issues

Zimmer Biomet is recalling Vital MIS Spinal Fixation system implants because reports indicate the extended tab screws may not lock adequately, causing the closure top to back out post-operatively.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a structural failure of a medical implant (screw backing out) which poses a risk of harm, but the source text does not report specific injuries or hospitalizations.

Plain-English summary

Zimmer Biomet, Inc. is recalling 3,235 units of Vital MIS Spinal Fixation system implants, specifically the CANN POLY EXT TAB 4.5X40MM (Item Number 810M4540). These non-cervical spinal fixation devices are intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation in skeletally mature and adolescent patients.

The recall was initiated due to reports of inadequate locking of the extended tab screws. Complaints were received indicating that the closure top backed out of the screw post-operatively. The affected units were distributed nationwide in the United States.

Healthcare providers and patients should contact Zimmer Biomet for instructions on how to proceed with the recall. The affected lot numbers are C20B0006, SBM108401, SBM118983, SBM121852, and SBM122267. The UDI for the product is (01)00887868324208.

The recalled product

Product
Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 4.5X40MM, Item Number 810M4540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is…
Affected units
3,235

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • UDI: (01)00887868324208
  • Lot Numbers: C20B0006 SBM108401 SBM118983 SBM121852 SBM122267

Distribution

Distributed nationwide across the United States.