RayStation Radiation Therapy Software Recalled for GPU Driver Error
RaySearch Laboratories is recalling RayStation software due to a potential calculation error in non-standard Citrix environments.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device software error that could lead to erroneous treatment calculations, representing a risk of harm where injury has not yet been reported.
Plain-English summary
RaySearch Laboratories AB is recalling 741 units of the RayStation Radiation Therapy Treatment Planning System, Models 4.5, 4.7, and 5.0. The recall is due to an error in NVIDIA GPU card drivers that can occur in specific software configurations.
In a non-standard Citrix environment where GPU settings are configured with a virtual Citrix graphics board, an erroneous calculation may be allowed. This software is used for treatment planning and analysis of radiation therapy, providing treatment unit set-up parameters and dose distribution estimates.
The affected units were distributed nationwide to 33 states and Washington D.C. Users should contact RaySearch Laboratories for further instructions regarding the software update or correction.
The recalled product
- Product
- Radiation Therapy Treatment Planning System, Model 4.5, 4.7, 5.0 RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during…
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Category
- Medical Device — Software
- Affected units
- 741
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (12)
- 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.3.13
- 4.7.4.4
- 4.7.5.4
- 5.0.0.37
- 5.0.1.11
- 5.0.2.35
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software error
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- HighPhilips Spectral CT 7500 Software Recall for Radiation Exposure Risk
FDA (Devices) · 2026-10-07
- HighBard EnCor EnCompass Breast Biopsy System Software Recall
FDA (Devices) · 2026-10-07
- ModerateBeckman Coulter AU480 Chemistry Analyzer Software Error Recall
FDA (Devices) · 2026-10-07
- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
Similar recalls
- HighRayStation Radiation Therapy Software Recalled for Potential Dose Errors
FDA (Devices) · 2014-04-09
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Errors
FDA (Devices) · 2016-12-14
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Accuracy Issues
FDA (Devices) · 2018-07-04
- HighRayStation software recall for potential incorrect treatment plan references
FDA (Devices) · 2020-12-02
- HighRayStation 5.0 Radiation Therapy Software Recalled for Dose Display Error
FDA (Devices) · 2016-12-14
- HighRayStation Radiation Therapy Software Recalled for Incorrect Biological Goal Evaluation
FDA (Devices) · 2016-07-27