RayStation Radiation Therapy Software Recalled for GPU Driver Error
RaySearch Laboratories is recalling RayStation software due to a potential calculation error in non-standard Citrix environments.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device software error that could lead to erroneous treatment calculations, representing a risk of harm where injury has not yet been reported.
Plain-English summary
RaySearch Laboratories AB is recalling 741 units of the RayStation Radiation Therapy Treatment Planning System, Models 4.5, 4.7, and 5.0. The recall is due to an error in NVIDIA GPU card drivers that can occur in specific software configurations.
In a non-standard Citrix environment where GPU settings are configured with a virtual Citrix graphics board, an erroneous calculation may be allowed. This software is used for treatment planning and analysis of radiation therapy, providing treatment unit set-up parameters and dose distribution estimates.
The affected units were distributed nationwide to 33 states and Washington D.C. Users should contact RaySearch Laboratories for further instructions regarding the software update or correction.
The recalled product
- Product
- Radiation Therapy Treatment Planning System, Model 4.5, 4.7, 5.0 RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during th
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Category
- Medical Device — Software
- Affected units
- 741
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (12)
- 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.3.13
- 4.7.4.4
- 4.7.5.4
- 5.0.0.37
- 5.0.1.11
- 5.0.2.35
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software error
See all →- HighDASH Rapid PCR Instrument Recalled Due to Software Issue Causing False Co-infection Results
FDA (Devices) · 2026-08-12
- HighQuantum Outback Powered Wheelchair Controller Recalled Due to Unexpected Braking Risk
FDA (Devices) · 2026-08-12
- HighQuantum Rehab Q6 Edge 3 Powered Wheelchair Controller Recalled for Unexpected Braking
FDA (Devices) · 2026-08-12
- HighQuantum Rehab R-TRAK Powered Wheelchair Controller Recalled Due to Software Issue
FDA (Devices) · 2026-08-12
- HighREMISOL Advance Software Version 2.5 Recalled for Incorrect Test Result Reporting
FDA (Devices) · 2026-08-12