RayStation 5.0 Radiation Therapy Software Recalled for Dose Display Error
RaySearch Laboratories is recalling RayStation 5.0 software due to a potential error in dose display on auxiliary CT images, which could result in incorrect treatment planning data.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk-of-harm product (radiation therapy planning software) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
RaySearch Laboratories AB is recalling 540 units of RayStation 5.0 Radiation Therapy Treatment Planning System software. The recall is due to a software error that may cause the display of computed dose on auxiliary CT images to be incorrect when multiple patient cases are used. This error can occur if a CT image set with the same Frame of Reference exists in another case, leading to incorrect maximum dose positions, dynamic isodose lines, or dose grid values.
The affected software versions are Build No. 5.0.0.37, 5.0.1.11, or 5.0.2.35, with only version 5.0.2.35 distributed in the United States. The software was distributed nationwide in the states of AZ, AR, CA, CT, DE, FL, GA, HI, IL, LA, MA, MD, MN, MI, MS, MO, NJ, NY, NC, OH, PA, RI, TX, and WA. The FDA has classified this as a Class II recall, indicating a reasonable probability that use of the product could cause temporary or medically reversible adverse health consequences.
Healthcare providers using RayStation 5.0 should verify their software version and contact RaySearch Laboratories for instructions on how to resolve the issue. The error affects the display of dose values in patient views, including the upper left corner display in 2D views, which may be incorrect or missing. Users should ensure that treatment plans are reviewed and approved by the intended user before administration, as per standard practice for this software.
The recalled product
- Product
- Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Affected units
- 540
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Build No: "5.0.0.37
- 5.0.1.11 or 5.0.2.35. Only 5.0.2.35 was distributed in the U.S.
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software error
See all →- HighDASH Rapid PCR Instrument Recalled Due to Software Issue Causing False Co-infection Results
FDA (Devices) · 2026-08-12
- HighQuantum Outback Powered Wheelchair Controller Recalled Due to Unexpected Braking Risk
FDA (Devices) · 2026-08-12
- HighQuantum Rehab Q6 Edge 3 Powered Wheelchair Controller Recalled for Unexpected Braking
FDA (Devices) · 2026-08-12
- HighQuantum Rehab R-TRAK Powered Wheelchair Controller Recalled Due to Software Issue
FDA (Devices) · 2026-08-12
- HighREMISOL Advance Software Version 2.5 Recalled for Incorrect Test Result Reporting
FDA (Devices) · 2026-08-12