Getinge Washer-Disinfector Accessory Recall for Potential Cart Unloading Hazard
Getinge is recalling 170 automated loading/unloading accessories for Washer-Disinfector 86 Series units due to a risk of disinfected goods falling to the floor.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product where injury has not yet been reported, fitting the criteria for a High severity score.
Plain-English summary
Getinge Disinfection Ab is recalling 170 units of the Getinge AGS, FSLC, FSUC, and RC accessories intended for the Washer-Disinfector 86 Series. These automated loading and unloading systems function as conveyor or transport mechanisms for wash carts and are considered accessories to the medical washer-disinfectors. The affected products were distributed worldwide, including in the United States and numerous international markets.
The recall is being issued because there is a potential for a cart containing disinfected goods to unload without the trolley being properly docked. If this occurs, the disinfected goods could fall to the floor, creating a risk of injury to nearby users and causing damage to the goods and equipment.
Healthcare facilities and users should identify the affected units by checking the serial numbers listed in the recall documentation. Users should contact Getinge Disinfection Ab for instructions on how to resolve the issue and ensure the safety of their equipment.
The recalled product
- Product
- Getinge AGS, FSLC, FSUC, RC accessory to Washer- Disinfector 86 Series UDI Info: 86 Series Washer Disinfector 07340153700246, 07340153700253, 07340153700260, 07340153700277 Getinge automated loading/unloading systems are intended to be used as an automatic conveyor/transpor
- Manufacturer
- Getinge Disinfection Ab
- Affected units
- 170
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- UDI Info: 86 Series Washer Disinfector 07340153700246
- 07340153700253
- 07340153700260
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Getinge Disinfection Ab under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Getinge Disinfection Ab
See all →- HighWasher disinfectors recalled over overheating risk from incorrect pressure calibration
FDA (Devices) · 2025-10-08
- ModerateWasher Disinfector Aquadis 56 Model 56M Recalled
FDA (Devices) · 2025-05-14
- ModerateWasher Disinfector Aquadis 56 Model 56A pump seal defect recall
FDA (Devices) · 2025-05-14
- HighGetinge Washer-Disinfector Accessory Recall Due to Unloading Hazard
FDA (Devices) · 2018-04-25
- LowGetinge 86 Series Washer/Disinfector Recalled for Remote Access Vulnerability
FDA (Devices) · 2015-08-26
Same hazard · falling object
See all →- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- SevereTerex Digger Derrick Trucks Recalled for Weld Defect
NHTSA · 2026-08-17
- SevereIsuzu NPR HD Liftgate Deck Pin Recall
NHTSA · 2026-08-06
- HighWabash Recalls 2025-2026 Dry Truck Bodies Due to Liftgate Pin Failure
NHTSA · 2026-07-14
- HighMarion Body Works Recalls Waltco Liftgates on Morgan Dry Vans
NHTSA · 2026-06-12