Washer Disinfector Aquadis 56 Model 56M Recalled
Getinge Disinfection Ab recalls the Washer Disinfector Aquadis 56 (Model 56M) because the frequency inverter is not programmed to specification, causing the circulation pump to start earlier than expected and wear the axial seal prematurely.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: This is an FDA Class II medical device recall. The source text does not report any illnesses, injuries, or failures; it describes a potential hazard (premature wear) with no documented adverse health consequences yet. This falls under the Moderate category as a precautionary recall involving a technical defect.
Plain-English summary
Getinge Disinfection Ab is recalling the Washer Disinfector Aquadis 56, Model Number 56M. The frequency inverter is not programmed by the supplier according to specification, which causes the circulation pump to start earlier than expected, resulting in accelerated wear and tear on the axial seal in the pump housing.
The affected units were distributed worldwide, including to the United States (79 units) and to 50 other countries. The recalled equipment is identified by UDI-DI 07340153710351 and includes specific serial numbers beginning with WAA.
If you own or operate an affected Washer Disinfector Aquadis 56 Model 56M, contact Getinge Disinfection Ab for instructions on remediation or replacement. Do not continue to operate equipment showing signs of pump seal failure.
The recalled product
- Product
- Washer Disinfector Aquadis 56; Model Number: 56M
- Manufacturer
- Getinge Disinfection Ab
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model: 56M
- UDI-DI: 07340153710351
- Serial Numbers: WAA105199
- WAA105251
- WAA105268
- WAA105357
- WAA105535
- WAA105700
- WAA105915
- WAA105972
- WAA105973
- WAA106079
- WAA106114
- WAA106135
- WAA106451
- WAA106471
- WAA106487
- WAA106505
- WAA107436
- WAA107492
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Getinge Disinfection Ab under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Getinge Disinfection Ab
See all →- HighWasher disinfectors recalled over overheating risk from incorrect pressure calibration
FDA (Devices) · 2025-10-08
- ModerateWasher Disinfector Aquadis 56 Model 56A pump seal defect recall
FDA (Devices) · 2025-05-14
- HighGetinge Washer-Disinfector Accessory Recall for Potential Cart Unloading Hazard
FDA (Devices) · 2018-04-25
- HighGetinge Washer-Disinfector Accessory Recall Due to Unloading Hazard
FDA (Devices) · 2018-04-25
- LowGetinge 86 Series Washer/Disinfector Recalled for Remote Access Vulnerability
FDA (Devices) · 2015-08-26
Same hazard · equipment malfunction
See all →- HighPhilips CT Systems Recalled for Unintended Radiation Performance Issues
FDA (Devices) · 2026-09-09
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- SevereHologic Selenia Dimensions Mammography Systems Loose Internal Bolts
FDA (Devices) · 2026-05-20
- SevereHalyard CATH LAB Kit Recalled for Syringe Adapter Disconnection Risk
FDA (Devices) · 2026-05-13