The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-1721-2025·Announced 2025-05-14

Washer Disinfector Aquadis 56 Model 56M Recalled

Getinge Disinfection Ab recalls the Washer Disinfector Aquadis 56 (Model 56M) because the frequency inverter is not programmed to specification, causing the circulation pump to start earlier than expected and wear the axial seal prematurely.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: This is an FDA Class II medical device recall. The source text does not report any illnesses, injuries, or failures; it describes a potential hazard (premature wear) with no documented adverse health consequences yet. This falls under the Moderate category as a precautionary recall involving a technical defect.

Plain-English summary

Getinge Disinfection Ab is recalling the Washer Disinfector Aquadis 56, Model Number 56M. The frequency inverter is not programmed by the supplier according to specification, which causes the circulation pump to start earlier than expected, resulting in accelerated wear and tear on the axial seal in the pump housing.

The affected units were distributed worldwide, including to the United States (79 units) and to 50 other countries. The recalled equipment is identified by UDI-DI 07340153710351 and includes specific serial numbers beginning with WAA.

If you own or operate an affected Washer Disinfector Aquadis 56 Model 56M, contact Getinge Disinfection Ab for instructions on remediation or replacement. Do not continue to operate equipment showing signs of pump seal failure.

The recalled product

Product
Washer Disinfector Aquadis 56; Model Number: 56M

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Model: 56M
  • UDI-DI: 07340153710351
  • Serial Numbers: WAA105199
  • WAA105251
  • WAA105268
  • WAA105357
  • WAA105535
  • WAA105700
  • WAA105915
  • WAA105972
  • WAA105973
  • WAA106079
  • WAA106114
  • WAA106135
  • WAA106451
  • WAA106471
  • WAA106487
  • WAA106505
  • WAA107436
  • WAA107492

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 2 recalled by Getinge Disinfection Ab under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 2 products in this recall →