The Recall Desk

FDA (Devices) recall action 96644

Getinge Disinfection Ab recalled 2 products on 2025-05-14

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-05-14
Critical
0
Units
—

Why these products were recalled

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

The agency’s own wording, quoted from the enforcement report.

1–2 of 2

  • ModerateFDA (Devices)··2025-05-14

    Washer Disinfector Aquadis 56 Model 56A pump seal defect recall

    Getinge Disinfection Ab is recalling Washer Disinfector Aquadis 56 (Model 56A) units due to a frequency inverter programming error that causes the circulation pump to start earlier than expected, leading to accelerated wear on the axial seal.

    Product
    Washer Disinfector Aquadis 56; Model Number: 56A
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-05-14

    Washer Disinfector Aquadis 56 Model 56M Recalled

    Getinge Disinfection Ab recalls the Washer Disinfector Aquadis 56 (Model 56M) because the frequency inverter is not programmed to specification, causing the circulation pump to start earlier than expected and wear the axial seal prematurely.

    Product
    Washer Disinfector Aquadis 56; Model Number: 56M
    Category
    Distribution
    0 states