Azurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
The deaeration hose in Philips Azurion X-ray tube cooling units may degrade over time and leak oil, affecting cooling performance and automatically triggering low-dose fluoroscopy mode.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II medical device recall affecting a critical cooling component of X-ray systems. Although no adverse health events are reported in the source, the hazard poses a risk of harm because degraded cooling performance and oil leakage could compromise system function and patient safety during diagnostic procedures.
Plain-English summary
Philips Medical Systems Nederland B.V. is recalling Azurion X-ray systems due to a defect in the deaeration hose found in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020. The hose may degrade over time, potentially resulting in oil leakage.
When oil leaks from the deaeration hose, the cooling performance of the X-ray tube is affected. If the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy mode and displays the message "Low load fluoroscopy flavor selected: Tube cooler problem." This condition cannot be resolved by restarting the system.
The recall affects 1,651 units worldwide: 592 in the United States and 1,359 internationally across numerous countries including Canada, Australia, France, Germany, Japan, and the United Kingdom. Affected facility operators should contact Philips for guidance on inspection and potential repair or replacement of the deaeration hose.
The recalled product
- Product
- Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079, 722224).
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1. Azurion 3 M12
- Model: 722063
- UDI: (01)00884838085275(21)64
- (01)00884838085275(21)56
- (01)00884838085275(21)77
- (01)00884838085275(21)82
- (01)00884838085275(21)47
- (01)00884838085275(21)80
- (01)00884838085275(21)102
- (01)00884838085275(21)117
- (01)00884838085275(21)2
- (01)00884838085275(21)99
- (01)00884838085275(21)44
- (01)00884838085275(21)3
- (01)00884838085275(21)62
- (01)00884838085275(21)61
- (01)00884838085275(21)22
- (01)00884838085275(21)29
- (01)00884838085275(21)40
- (01)00884838085275(21)70
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
See all →- HighPhilips Azurion Rystem X-Ray System Software Recall for Image Quality Issues
FDA (Devices) · 2026-08-19
- HighPhilips Allura Xper FD Fluoroscopic X-Ray System Recalled for Firmware Issue
FDA (Devices) · 2026-08-19
- HighPhilips Azurion Fluoroscopy System Recall for Software Image Quality Issues
FDA (Devices) · 2026-08-19
- HighPhilips Fluoroscopy System Recall Over Hard Disk Drive Issue
FDA (Devices) · 2026-08-12
- SeverePhilips Allura X-Ray Systems Recalled Over Unauthorized Hard Disk Drives
FDA (Devices) · 2026-08-12
Same hazard · oil leakage
See all →- ModeratePhilips Allura X-ray Systems Deaeration Hose Degradation Recall
FDA (Devices) · 2026-05-27
- HighIntegris-Allura X-ray systems with degraded deaeration hoses recalled
FDA (Devices) · 2026-05-27
- HighPhilips Ingenuity TF PET/CT Systems Recalled for X-Ray Tube Oil Leakage
FDA (Devices) · 2018-07-04
- HighPhilips Ingenuity Core128/Elite X-Ray Tube Recall for Oil Leakage
FDA (Devices) · 2018-07-04
- HighPhilips Brilliance Big Bore Radiology Systems Recalled for X-Ray Tube Defect
FDA (Devices) · 2018-07-04
Similar recalls
- HighPhilips Azurion X-ray Systems: Cable Hose Carrier Detachment Risk
FDA (Devices) · 2024-09-25
- ModeratePhilips Azurion X-Ray System Cable Hose Carriers May Loosen or Break
FDA (Devices) · 2026-05-20
- ModeratePhilips Recalls Azurion X-ray Systems Due to Software Defect
FDA (Devices) · 2019-08-07
- ModeratePhilips Azurion X-ray systems recalled for cold restart delays
FDA (Devices) · 2019-08-14
- HighPhilips Azurion X-ray System: Potential Loss of Imaging Functionality
FDA (Devices) · 2024-08-14
- HighPhilips Azurion system hard drives may degrade and lose imaging
FDA (Devices) · 2026-06-03