Intuitive Surgical Recalls da Vinci Xi Systems Due to Sensor Fault
Intuitive Surgical is recalling 167 da Vinci Xi Surgical Systems worldwide due to a fault causing distal SUJ failures and position sensor mismatches.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a high-risk medical device (surgical robot) with a mechanical fault that could compromise safety, but the source text does not report any injuries or illnesses, fitting the 'High' criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Intuitive Surgical, Inc. is recalling 167 da Vinci Xi Surgical Systems (Model IS4000) with material numbers 380647-20 and 380647-22. The recall is due to a fault that causes distal SUJ (DSUJ) failures, which indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.
The affected systems were distributed worldwide, including the United States (nationwide) and countries such as Australia, Belgium, China, France, Germany, India, Israel, Italy, Norway, South Korea, Spain, Sweden, Switzerland, Turkey, and the United Kingdom. The source text does not specify the number of reported injuries or illnesses associated with this defect.
Healthcare facilities should contact Intuitive Surgical, Inc. for instructions on how to address the defect. The source text does not provide specific consumer action steps beyond the recall notification.
The recalled product
- Product
- da Vinci Xi Surgical System, IS4000; ASSY,USM,IS4000, Material number: 380647-20 and 380647-22 The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic…
- Manufacturer
- Intuitive Surgical, Inc.
- Category
- Medical Device — Surgical Robot
- Affected units
- 167
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Material number: 380647-20
- Serial numbers: 1311140430
- 1311140586
- 1311140058
- 1311140392. 1311140277
- 1311140307
- 1311140094
- 1311140100
- 1311140569
- 1311140102
- 1311140421
- 1311140092
- 1311140009
- 1311140070
- 1311140267
- 1311140333
- 1311140492
- 1311140391
- 1311140057
- 1311140294
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Intuitive Surgical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Intuitive Surgical, Inc.
See all →- Moderateda Vinci SP Instrument Arm Drape may detach during surgical procedures
FDA (Devices) · 2026-07-22
- SevereDa Vinci surgical system operative components with E-brake retainer pin defect
FDA (Devices) · 2026-07-01
- HighDa Vinci 5 tower may unintentionally activate the infrared handheld camera
FDA (Devices) · 2026-06-10
- Highda Vinci X and Xi Surgical Systems Recalled for Manipulator Arm Screw Breakage
FDA (Devices) · 2026-05-13
- SevereIntuitive 8 mm SureForm 30 Gray Reloads Incomplete Staple Defect
FDA (Devices) · 2026-05-13
Same hazard · mechanical failure
See all →- ModerateLeMaitre TufTex Over-the-Wire Embolectomy Catheter Recalled for Balloon Failure
FDA (Devices) · 2026-10-14
- SevereCanon Medical Alphenix INFX-8000C/B5 Recalled for Tabletop Shift Risk
FDA (Devices) · 2026-10-14
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
Similar recalls
- HighIntuitive Surgical Recalls da Vinci Xi Patient Side Carts Due to Sensor Failure
FDA (Devices) · 2014-10-08
- HighIntuitive Surgical Recalls da Vinci Xi Systems Due to Software Anomaly
FDA (Devices) · 2016-11-02
- HighIntuitive Surgical Recalls da Vinci Xi Surgical Systems Due to Brake Screw Issue
FDA (Devices) · 2017-07-05
- HighIntuitive Surgical Recalls da Vinci X Systems with Defective Arm Controller Boards
FDA (Devices) · 2018-07-11
- HighIntuitive Surgical Recalls da Vinci Xi Systems Due to Defective Capacitors
FDA (Devices) · 2020-02-12
- SevereIntuitive Surgical Recalls da Vinci Xi Systems Due to Arm Faults
FDA (Devices) · 2020-05-20