The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2630-2017·Announced 2017-07-05

Intuitive Surgical Recalls da Vinci Xi Surgical Systems Due to Brake Screw Issue

Intuitive Surgical is recalling 10 da Vinci Xi Surgical Systems because brake screws on the patient side carts may be under-torqued.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a mechanical defect in a surgical device that could compromise safety during procedures, fitting the rubric for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Intuitive Surgical, Inc. is recalling 10 units of the da Vinci Xi Surgical System (IS4000). The recall is due to a defect where the patient side carts (PSC) were shipped with potentially under-torqued brake screws in a section of the system arms.

The da Vinci Xi Surgical System is designed to assist surgeons in controlling instruments during endoscopic surgery. The affected units were distributed worldwide, including in the United States (Texas, New York, Florida, Illinois, and Ohio) and in France, Ireland, Japan, and Taiwan.

Healthcare facilities should contact Intuitive Surgical for instructions on how to address the defect. The specific serial numbers for the affected units are SK0822, SK1154, SK1001, SK1170, SK1040, SK1176, SK1082, SK1180, SK1146, and SK1190.

The recalled product

Product
da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in the accurate control of surgical instruments such as endoscopes, scissors, forceps, and retractors during endoscopic surgery. The system has four universal…
Affected units
10

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (10)

  • Serial Numbers: SK0822
  • SK1154
  • SK1001
  • SK1170
  • SK1040
  • SK1176
  • SK1082
  • SK1180
  • SK1146
  • SK1190

Distribution

Distributed nationwide across the United States.

The notice also names 5 states specifically: