Philips IntelliSpace Portal Software Recall for URL PACS Integration Issue
Philips is recalling IntelliSpace Portal software version 4.0.2 due to a display error when using URL PACS integration. The issue causes the Review monitor to display incorrect patient studies.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves the potential for clinicians to review incorrect patient images, which is a risk-of-harm product issue where injury has not yet been reported.
Plain-English summary
Philips Medical Systems (Cleveland) Inc. is recalling 313 units of IntelliSpace Portal (ISP) software, specifically version 4.0.2, due to a defect in the URL PACS integration configuration. The IntelliSpace Portal is a multimodality thin-client applications server that delivers diagnostic viewing and clinical applications to healthcare enterprises.
The recall was initiated after Philips was notified that when loading a study to an analysis application using the URL PACS integration, the Review monitor may fail to update correctly. Specifically, if another study is opened by a review application, the Analysis monitor updates with the current patient's study, but the Review monitor continues to display the previous study. This discrepancy could lead to the review of incorrect patient data.
The affected units were distributed worldwide, including nationwide distribution in the United States and in numerous international countries. Healthcare facilities using this specific software version with URL PACS integration should verify their configuration and contact Philips for corrective action.
The recalled product
- Product
- Intellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is a multimodality (CT, NM, and MR) thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. IntelliSpace Portal (ISP) is a…
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Hazard
- Affected units
- 313
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model #881001
- Serial #s: 79916
- 79932
- 79933
- 79935
- 79936
- 79937
- 79939
- 79940
- 79943
- 80071
- 80914
- 80916
- 80925
- 80926
- 80928
- 80932
- 80939
- 80963
- 80965
Distribution
Related recalls
Same manufacturer · Philips Medical Systems (Cleveland) Inc
See all →- HighRadiation planning software recalled over image errors in three patient orientations
FDA (Devices) · 2025-09-24
- HighPinnacle 3 radiation planning software may miscalculate stopping power ratio
FDA (Devices) · 2025-08-13
- HighPinnacle3 Radiation Therapy Planning System dose-calculation software error
FDA (Devices) · 2022-09-21
- SeverePhilips Forte Gamma Camera System Recalled Due to Detector Falling Hazard
FDA (Devices) · 2019-11-13
- SeverePhilips Recalls Forte Gamma Camera Systems Due to Detector Fall Risk
FDA (Devices) · 2019-11-13
Same hazard · incorrect data display
See all →- HighInvivo DynaCAD Software Recalled for Incorrect Pharmacokinetic Map Display
FDA (Devices) · 2018-04-11
- HighToshiba APLIO 300 Ultrasound Systems Recalled for Incorrect Velocity Trace Display
FDA (Devices) · 2017-07-19
- HighMerge PACS Software Recalled for Potential Incorrect Patient Image Display
FDA (Devices) · 2017-02-15
- HighSiemens Syngo Dynamics PACS Recalled for Patient Demographics Display Error
FDA (Devices) · 2014-05-21
- HighPhilips CT System Software Recall for Incorrect Value Display
FDA (Devices) · 2013-03-13
Similar recalls
- ModeratePhilips Recalls Intellispace Portal Software Due to Missing Warning Messages
FDA (Devices) · 2012-07-18
- HighPhilips Intellispace Portal software recalled for tumor measurement errors
FDA (Devices) · 2012-07-18
- ModeratePhilips Recalls IntelliSpace Portal Software Due to MUGA Scan Display Error
FDA (Devices) · 2014-08-20
- ModeratePhilips Medical Systems Recalls IntelliSpace Portal and Extended Brillance Workspace Software
FDA (Devices) · 2012-07-18
- HighPhilips IntelliSpace Portal Software Recall for SPECT Reconstruction Errors
FDA (Devices) · 2014-06-11
- HighPhilips IntelliSpace PACS 4.4 Software Recall for Data Integrity Issues
FDA (Devices) · 2014-08-27