Invivo DynaCAD Software Recalled for Incorrect Pharmacokinetic Map Display
Invivo Corporation is recalling DynaCAD software versions 3.4 and 3.5 because pharmacokinetic and ADC maps may display incorrectly on remote clients, potentially leading to underestimated values.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves incorrect data display in medical software which could lead to misinterpretation of diagnostic results, fitting the criterion for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Invivo Corporation is recalling DynaCAD software, a post-processing package used for analyzing MRI datasets. The recall affects software versions 3.4 and 3.5. The product consists of an MR Analysis Server and viewer workstation software, including modules for breast, prostate, and other MR analyses.
The recall is due to a defect identified after a software update from version 3.3 to 3.5. The defect causes pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client computers. This error could result in the visual underestimation of calculated Ktrans, Kep, and iAUGC values. Additionally, the defect impacts the on-the-fly calculation of Apparent Diffusion Coefficient (ADC) maps; if computed by DynaCAD, these values and colors will also display incorrectly on remote clients.
Users are advised to discontinue the use of Ktrans, Kep, and iAUGC colormaps when assessing studies from remotely connected DynaCAD client computers. Users should also refrain from using colormaps and values derived from ADC maps calculated by DynaCAD. Instead, users should utilize ADC maps originating natively from the MRI system, as these values are unaffected by the defect. Invivo Corporation will provide a v4.0 software update to correct the defect at no charge to the user.
The recalled product
- Product
- DynaCad software consist of an MR Analysis Server software and the viewer workstation software. The MR Analysis software consists of DynaCad Breast, DynaCad Prostate, and DynaCad Advanced PK for other MR analyses modules. Product Usage: Intended to be used as a post…
- Manufacturer
- Invivo Corporation
- Category
- Medical Device — Software
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Invivo Corporation
See all →- LowInvivo MR Coils Recalled for Incorrect IPX1 Water Ingress Rating Label
FDA (Devices) · 2020-10-28
- HighInvivo Recalls MRI Monitor Adapters Due to Insufficient Shielding
FDA (Devices) · 2020-03-11
- HighInvivo Xper Flex Cardio Physiomonitoring System Software Recall
FDA (Devices) · 2018-11-21
- HighInvivo Xper Flex Cardio Physiomonitoring System Recalled for Firmware Error
FDA (Devices) · 2018-09-26
- ModerateInvivo Xper Flex Cardio Physiomonitoring System Recalled for Software Defect
FDA (Devices) · 2018-06-13
Same hazard · incorrect data display
See all →- HighToshiba APLIO 300 Ultrasound Systems Recalled for Incorrect Velocity Trace Display
FDA (Devices) · 2017-07-19
- HighMerge PACS Software Recalled for Potential Incorrect Patient Image Display
FDA (Devices) · 2017-02-15
- HighSiemens Syngo Dynamics PACS Recalled for Patient Demographics Display Error
FDA (Devices) · 2014-05-21
- HighPhilips IntelliSpace Portal Software Recall for URL PACS Integration Issue
FDA (Devices) · 2013-04-24
- HighPhilips CT System Software Recall for Incorrect Value Display
FDA (Devices) · 2013-03-13
Similar recalls
- HighiCAD Recalls DynaLOC Breast Interventional Planning Software Versions 3.1 and 3.2
FDA (Devices) · 2015-03-11
- HighBrainlab Patient Data Manager Software Recalled for Image Display Errors
FDA (Devices) · 2015-11-11
- ModerateDICOM GRID Recalls Ambra PACS Software Due to Image Display Error
FDA (Devices) · 2017-01-25
- HighBrainlab ExacTrac 6.0 Patient Positioning System Recalled for Display Error
FDA (Devices) · 2015-09-02
- HighHitachi VidiStar PACS Software Recalled for Potential Image Mix-Up
FDA (Devices) · 2018-09-12
- HighElekta Monaco Software Recall for Incorrect Patient Shift Directions
FDA (Devices) · 2012-07-18